front 1 In many cases, digital fluoroscopy eliminates which of the following? a. The need for a radiographer b. Postprocedure “overhead” images c. Image acquisition d. Accurate positioning | back 1 B. Postprocedure “overhead” images -Most views required by the radiologist are acquired digitally during fluoroscopy. |
front 2 What is the smallest particle of an element that retains the characteristics of the element? a. Mole b. Atom c. Molecule d. Quark | back 2 B. Atom -The atom may be broken down into subatomic particles but they would not have the characteristics of the element. |
front 3 In digital fluoroscopy, the image must be turned into digital form by what device? a. Digital-to-analog converter b. Flux capacitor c. Analog-to-digital converter d. DVD-ROM | back 3 C. Analog-to-digital converter -The information acquired must be converted from its analog form to digital for further manipulation. |
front 4 X-rays travel as bundles of energy called: a. Energy waves b. Phasers c. Electromagnetic bursts d. Photons | back 4 D. Photons -Photons may also be called quanta. |
front 5 What is atomic mass? a. The number of protons plus the number of neutrons b. The number of photons c. The number of electrons d. The number of protons plus the number of electrons | back 5 A. The number of protons plus the number of neutrons -Atomic number is the number of protons. Protons and neutrons are contained in the atomic nucleus. |
front 6 What device may be used to ensure consistency of radiographic quality from one exposure to the next? a. Electronic timer b. Step-up transformer c. Automatic exposure control d. High-frequency generator | back 6 C. Automatic exposure control -The AEC is calibrated to provide the images most preferred by the radiologists. |
front 7 When a predetermined level of ionization is reached in the ionization chamber, what does the machine do? a. The unit shuts off as a result of a malfunction b. The maximum allowable time has been reached c. The highest allowable dose to the patient has been reached d. The exposure is terminated | back 7 D. The exposure is terminated -Hence, the exposure time when using an AEC is very dependent on the kVp. |
front 8 What type of x-ray machine uses a continually decreasing mA for the shortest times possible? a. Ionization chamber b. Portable c. C-arm d. Falling load generator | back 8 D. Falling load generator -The falling load generator makes use of the maximum heat storage capacity at every mA and time combination. This allows for the shortest exposure time possible. |
front 9 What type of current is required for proper operation of the x-ray tube? a. Direct b. Falling load c. Alternating d. Fluctuating | back 9 A. Direct -This necessitates the use of a rectifier to change AC to DC. |
front 10 The law stating that the outer shell of an atom can contain no more than eight electrons is called: a. Ohm’s law b. Octet rule c. Octagon rule d. Electron binding energy | back 10 B. Octet rule -The octet rule states that no more than eight electrons can be in the outer shell. It does not stipulate that there must be eight electrons in the outer shell. |
front 11 Examples of particulate radiation are: a. X-rays, gamma rays, and cosmic rays b. Helium nuclei and beta particles c. Electrons, protons, and meteorites d. X-rays and quarks | back 11 B. Helium nuclei and beta particles -Particulate radiations are highly ionizing. |
front 12 Electromagnetic radiation travels: a. In waves along a straight path b. In circles c. Back and forth d. As electrons in waves along a straight path | back 12 A. In waves along a straight path -Because it has mass, particulate radiation may travel along different paths. |
front 13 The latest digital fluoroscopic units use: a. Nonionizing radiation b. CR cassettes c. Conventional image intensifiers d. Flat panel plates similar to DR | back 13 D. Flat panel plates similar to DR |
front 14 At what speed do x-rays travel? a. The speed of light—186,000 miles per hour b. The speed of the incident electrons—93,000 miles per second c. The speed of light—186,000 miles per second d. Infinite speed | back 14 C. The speed of light—186,000 miles per second -Do not be fooled by choices that use different units of measurement. Light speed is expressed in miles per second, not miles per hour. |
front 15 Waves of radiation are called: a. Sine waves b. Strong waves c. Signal waves d. Current waves | back 15 A. Sine waves |
front 16 Wavelength is defined as the distance from: a. The x-ray tube to the patient b. The cathode to the anode c. The bottom of a wave to the top d. Peak to peak of the wave | back 16 D. Peak to peak of the wave -Wavelength can be measured from crest to crest or trough to trough. |
front 17 Frequency is defined as: a. Synonymous with wavelength b. The number of waves passing a point per unit time c. The number of waves striking the patient d. The number of exposures needed during an exam | back 17 B. The number of waves passing a point per unit time -Higher-frequency waves have shorter wavelengths; lower-frequency waves have longer wavelengths. |
front 18 The speed of x-rays is based on: a. mAs b. kVp c. Size of the patient d. The fact that they are a form of electromagnetic radiation | back 18 D. The fact that they are a form of electromagnetic radiation. -The speed of x-ray travel is constant, regardless of the factors used to produce them. |
front 19 Wavelength and frequency are: a. Directly proportional to each other b. Inversely proportional to each other c. Inversely proportional to the square of their distance d. Unrelated to each other | back 19 B. Inversely proportional to each other |
front 20 The x-ray beam changes as it travels through the patient by a process called: a. Filtration b. Attenuation c. Electrification d. Annihilation | back 20 B. Attenuation -Attenuation means that the x-rays may be absorbed or scattered |
front 21 The intensity of radiation is inversely proportional to the square of the distance between the source of radiation and the person receiving it. This describes the: a. Square law b. Reciprocity law c. Inverse square law d. Octet law | back 21 C. Inverse square law -Be certain to review the inverse square law; always remember the inverse square relationship. The resulting answer cannot be double or half; it will be ¼ or 4 times. |
front 22 As radiation strikes matter: a. The energy of the rays is destroyed b. The energy of the rays is increased as they acquire the energy of the atoms c. The energy is transferred to the atoms according to the law of conservation of energy d. The energy is converted to matter according to Einstein’s theory | back 22 C. The energy is transferred to the atoms according to the law of conservation of energy -Energy cannot be created or destroyed, only changed in form. The energy is merely transferred. |
front 23 Choose all of the following statements that are accurate regarding electrostatic charges. a. Electrostatics is the study of electric charges at rest b. The movement of electrons from one object to another is called ionization c. Like charges attract, and unlike charges repel d. Electrostatic charges concentrate on a conductor in the area of greatest curvature e. Friction, contact, and induction are methods of ionization | back 23 A. Electrostatics is the study of electric charges at rest D. Electrostatic charges concentrate on a conductor in the area of greatest curvature |
front 24 Choose all of the following that are accurate (choose 6): a. A magnetic field always surrounds an electrical charge in motion b. Current flows back and forth in AC c. Current flows in one direction in DC d. The volt is the unit of electrical current e. A conductor allows the flow of electrons f. The ampere is the unit of electromotive force g. The volt is the unit of potential difference h. The path of electrical current is called the circuit | back 24 A. A magnetic field always surrounds an electrical charge in motion B. Current flows back and forth in AC C. Current flows in one direction in DC E. A conductor allows the flow of electrons G. The volt is the unit of potential difference H. The path of electrical current is called the circuit |
front 25 Electromagnetic induction is the process of causing an electrical current to flow in a conductor: a. When it is placed in contact with another conductor b. When it is placed in contact with an insulator c. When it is placed in contact with a superconductor d. When it is placed in the magnetic field of another conductor | back 25 D. When it is placed in the magnetic field of another conductor -Electromagnetic induction does not require that two conductors touch each other. |
front 26 The two types of electromagnetic induction are: a. Autoinduction and mutual induction b. Self-induction and mutual induction c. Generated induction and self-induction d. Current induction and voltage induction | back 26 B. Self-induction and mutual induction -Self-induction occurs in the autotransformer; mutual induction occurs in the step-up and stepdown transformers. |
front 27 Self-induction is used in the operation of what device? a. Step-up transformer b. Autotransformer c. Step-down transformer d. Electronic timer | back 27 B. Autotransformer -Mutual induction occurs in the step-up and stepdown transformers. |
front 28 The strength of the magnetic fields in a transformer is increased by: a. Coiling the wires and placing them in adjoining machines b. Coiling the wires and letting their magnetic fields overlap c. Keeping the wires very straight, increasing their effectiveness d. Replacing the wires with diodes | back 28 B. Coiling the wires and letting their magnetic fields overlap -This intensifies the magnetic fields, making the transformers more efficient. |
front 29 Electricity is supplied to the imaging department by a: a. Motor b. Rectifier c. Generator d. Voltmeter | back 29 C. Generator -A generator at the power company converts mechanical energy to electrical energy. A motor converts electrical energy to mechanical energy. |
front 30 The electricity provided to the radiology department is: a. 110 Hz or 220 Hz DC b. 110 Hz or 220 Hz DC c. 60 Hz AC d. 60 Hz DC | back 30 C. 60 Hz AC -60 Hertz means 60 cycles per second. The voltage may be 110 or 220. |
front 31 The electricity provided to the radiology department operates at: a. 120 pulses per second b. 60 pulses per second c. 110 pulses per second d. 220 pulses per second | back 31 A. 120 pulses per second -60 cycles per second results in 120 pulses per second because it cycles back and forth 120 times per second. |
front 32 High-frequency power: a. Is less effective than single-phase power b. Has almost no ripple c. Has more ripple than three-phase power d. Is yet unproved | back 32 B. Has almost no ripple -Such a low ripple means higher average photon energy. High frequency is very efficient. |
front 33 The primary advantage of three-phase power is that: a. Voltage drops to zero only 6 times per second b. Voltage drops to zero only 12 times per second c. Voltage never drops to zero d. Voltage is always at peak value | back 33 C. Voltage never drops to zero -Three-phase power produces higher average photon energy than single-phase power. |
front 34 A variable transformer that is used to select kVp for the x-ray circuit is the: a. Step-up transformer b. Autotransformer c. Step-down transformer d. Rectifier | back 34 B. Autotransformer -The autotransformer is a variable transformer that operates on the principle of self-induction. |
front 35 A transformer that has more turns in the secondary coil than in the primary coil is called a: a. Step-up transformer b. Solenoid c. Step-down transformer d. Filament transformer | back 35 A. Step-up transformer -Voltage is stepped up while current is stepped down in the same proportion. A transformer is named for what it does to voltage. |
front 36 What is the transformer used to boost voltage to kilovoltage levels called? a. Autotransformer b. Step-down transformer c. Step-up transformer d. Low-voltage transformer | back 36 C. Step-up transformer -The voltage is varied at the autotransformer, then boosted to kilovoltage levels by the step-up transformer. |
front 37 Voltage coming to the x-ray machine is kept constant through the use of a(n): a. Autotransformer b. Step-down transformer c. Rectifier d. Line voltage compensator | back 37 D. Line voltage compensator -Older x-ray machines require that the line voltage compensator be adjusted manually, whereas newer equipment lets the machine do the adjustment itself. |
front 38 A step-down transformer: a. Steps down voltage b. Steps down current c. Steps up voltage d. Steps up resistance | back 38 A. Steps down voltage -A transformer is named for what it does to voltage. |
front 39 Where does thermionic emission occur? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 39 C. Cathode -Thermionic emission is the boiling off of electrons from the filament wire. |
front 40 Which of the following devices is prereading? a. Step-down transformer b. Rectifier c. Timer d. kVp meter | back 40 D. kVp meter -The kVp meter indicates what the voltage will be after being stepped up. Hence it is prereading. |
front 41 Which device reduces voltage and provides current to produce an electron cloud or space charge at the filament? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 41 A. kVp meter -The step-down transformer is located in the filament circuit. |
front 42 Which device is electronic, with a setting as low as 0.001? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 42 D. Timer -The timer is accurate as low as 1⁄1000 of a second (which is 1 ms or 0.001 second). |
front 43 What changes AC to DC? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 43 B. Rectifier -The rectifier changes incoming alternating current from the power company into direct current for use by the x-ray tube. |
front 44 Which of the following is surrounded by a negatively charged focusing cup? a. Step-down transformer b. Rectifier c. Filament d. Timer | back 44 C. Filament -The filament and focusing cup are part of the cathode assembly. The focusing cup has a negative charge at the moment of exposure so as to repel the electrons from the cathode. |
front 45 Which of the following is composed of solid-state silicon-based diodes? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 45 B. Rectifier -The rectifier changes AC to DC. |
front 46 What is located in the x-ray circuit between the high voltage transformer and the x-ray tube? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 46 B. Rectifier -At this point, AC is changed to DC so that current is flowing in only one direction through the x-ray tube. |
front 47 Which of the following measures tube current? a. Anode b. mA meter c. Ionization chamber d. Falling load generator | back 47 B. mA meter -Current is the flow of electrons, as measured in milliamperes (mA). Hence the mA meter provides the reading. |
front 48 What device spins at 10,000-12,000 rpm? a. Anode b. mA meter c. Ionization chamber d. Step-up transformer | back 48 A. Anode -High-speed anodes spin from 10,000 to 12,000 rotations per minute (rpm). |
front 49 What device uses maximum heat storage ability of the tube to deliver mAs? a. Anode b. mA meter c. Ionization chamber d. Falling load generator | back 49 D. Falling load generator -This type of generator allows extremely short exposure times. |
front 50 What is the source of bremsstrahlung and characteristic rays? a. Anode b. mA meter c. Falling load generator d. Step-up transformer | back 50 A. Anode -These are the x-rays produced by two types of interactions that occur between incident electrons from the cathode and the tungsten atoms of the anode. |
front 51 What device increases voltage approximately 500 times? a. Anode b. mA meter c. Ionization chamber d. Step-up transformer | back 51 D. Step-up transformer -This is where voltage is boosted to kilovoltage levels. |
front 52 What is the most commonly used AEC? a. Phototimer b. mA meter c. Ionization chamber d. Falling load generator | back 52 C. Ionization chamber -The ionization chamber is located between the patient and the image receptor. Phototimers are seldom used. |
front 53 What device always delivers the shortest exposure time possible? a. Anode b. mA meter c. Ionization chamber d. Falling load generator | back 53 D. Falling load generator -Coupled with an electronic timer, this results in times as short as 1⁄1000 second. |
front 54 What device is turned by a rotor? a. Anode b. mA meter c. Ionization chamber d. Step-up transformer | back 54 A. Anode -Pressing the rotor button activates a motor that turns the anode. It also begins thermionic emission at the cathode. |
front 55 Which of the following is located between the patient and the image receptor? a. Anode b. mA meter c. Ionization chamber d. Falling load generator | back 55 C. Ionization chamber -The ionization chamber is a wafer-thin chamber containing gas that is ionized by the x-rays passing through it. At a predetermined level of ionization, the exposure is terminated. |
front 56 The filament is kept warm by: a. A standby current from the time the x-ray machine is turned on b. Insulating oil c. Lead housing d. Current produced only during exposures | back 56 A. A standby current from the time the x-ray machine is turned on -The machine keeps the filament warm between exposures. |
front 57 Activating the rotor: a. Accelerates a procedure b. Reduces tube life c. Keeps it oiled d. Keeps the cooling fan activated | back 57 B. Reduces tube life -Each time the rotor is activated, thermionic emission begins to occur at the filament. This heating of the filament causes tungsten to evaporate, thus slowly deteriorating the filament. |
front 58 When making an exposure, the radiographer should: a. Hold the rotor for several seconds before pressing “expose” b. Activate the rotor and exposure switch in one continuous motion c. Begin the rotor while a student is still positioning the patient d. Activate the rotor while estimating how fast the anode is spinning | back 58 B. Activate the rotor and exposure switch in one continuous motion -Choice C is incorrect, although many first-year students experience this activity when technologists try to hurry them from the room! Choices A and D are incorrect because unnecessarily heating the filament actually reduces x-ray tube life. |
front 59 The process of thermionic emission causes: a. Electrons to boil off the anode b. The anode to spin c. The cathode to cool quickly d. Electrons to boil off the filament | back 59 D. Electrons to boil off the filament -This causes a space charge or electron cloud to form around the filament. |
front 60 The force with which the electron stream passes from cathode to anode is a result of _____ passing through the x-ray tube. a. Current b. Kilovoltage c. Tungsten d. Heat | back 60 B. Kilovoltage -Voltage is the force behind the movement of electrons in a circuit. |
front 61 Heat is produced in the x-ray tube as: a. Electrons break apart while striking the anode b. Electrons interact with the target material c. The anode stops spinning d. The rectifier operates | back 61 B. Electrons interact with the target material -This is primarily a result of bremsstrahlung interactions at the anode. |
front 62 Most of the energy conversion in the x-ray tube produces: a. X-rays b. Light c. Heat d. Current | back 62 C. Heat -X-rays and heat are produced at the anode. This is the result of a massive energy conversion in which more than 99% of the incident electrons’ kinetic energy is converted to heat energy, while less than 1% is converted to x-ray energy. This is a clear example of the law of conservation of energy. No light is produced in the x-ray tube. |
front 63 X-rays are produced as incident electrons are slowed going past the nuclei of target atoms by a process called: a. Classical b. Photoelectric c. Bremsstrahlung d. Compton | back 63 C. Bremsstrahlung -As incident electrons fly past the atomic nuclei, they are slowed, causing the production of heat and x-rays by bremsstrahlung. X-rays are also produced by the characteristic interaction. |
front 64 X-rays are produced as incident electrons collide with inner shell electrons in target atoms by a process called: a. Characteristic b. Photoelectric c. Bremsstrahlung d. Compton | back 64 A. Characteristic -As incident electrons dislodge K-shell electrons in the target material, outer shell electrons fall inward to fill the holes. This causes the release of energy in the form of x-rays, with energy characteristic of the difference in energy levels between the K-shell and L-shell. |
front 65 What percentage of energy in the x-ray tube is converted to x-rays? a. 99% b. 75% c. 1% d. 100% | back 65 C. 1% -More than 99% of the energy is converted to heat. |
front 66 The x-ray beam is: a. Heterogeneous—all rays possess the same energy b. Homogeneous—all rays possess the same energy c. Monoenergetic—all energies correspond to the kVp d. Heterogeneous or polyenergetic—consisting of many different energies (wavelengths) | back 66 D. Heterogeneous or polyenergetic—consisting of many different energies (wavelengths) -Choice A is correct in stating that the x-ray beam is heterogeneous, but the second half of choice A is incorrect. |
front 67 The x-ray emission spectrum consists of: a. Brems and characteristic rays b. Discrete spectrum (produced by brems rays) and continuous spectrum (produced by characteristic rays) c. Discrete spectrum (produced by characteristic rays) and continuous spectrum (produced by brems rays) d. X-rays and electrons, both part of the electromagnetic spectrum | back 67 C. Discrete spectrum (produced by characteristic rays) and continuous spectrum (produced by brems rays) -We do not often discuss the x-ray beam in this detail, but be sure to review discrete and continuous in Chapter 3. |
front 68 The primary purpose of filtration is: a. Radiation protection b. Removal of short-wavelength (soft) rays c. Hardening the beam for imaging d. Removal of long-wavelength (hard) rays | back 68 A. Radiation protection -Choice B is incorrect because soft rays have long wavelengths. Choice C is true, but it is not the primary purpose of filtration.Choice D is incorrect because long-wavelength rays are called soft rays. |
front 69 The amount of material needed to reduce the intensity of the beam by 1/10 is called: a. Half-value layer b. Tenth-value layer c. Total filtration d. Inherent filtration | back 69 B. Tenth-value layer -We more commonly discuss beam quality using the term half-value layer. Regardless of the fraction, it is always the amount of material needed to reduce beam intensity to that level. |
front 70 Which of the following statements regarding filtration is true? a. Total filtration must not be less than 2-mm aluminum equivalent b. Total filtration must remove all soft rays from the beam c. Total filtration (added + compensating) must not be less than 2.5-mm aluminum equivalent d. Total filtration (not less than 2.5-mm aluminum equivalent) = inherent filtration (glass envelope, tube housing, oil) + added filtration (aluminum) | back 70 D. Total filtration (not less than 2.5-mm aluminum equivalent) = inherent filtration (glass envelope, tube housing, oil) + added filtration (aluminum) -These are detailed statements attempting to define the amount of filtration required. D is the most complete and accurate of all the choices. A is incorrect as to amount, B is incorrect because we cannot remove all soft rays from the beam, and C is incorrect because compensating filtration is added only for certain exams. |
front 71 Calculating heat units for three-phase, 12-pulse equipment requires the use of _____ as a constant; calculating heat units for single-phase equipment requires the use of _____ as a constant; calculating heat units for three-phase, 6-pulse equipment requires the use of _____ as a constant. a. 1.0, 1.35, 1.41 b. 1.35, 1.0, 1.41 c. Calculating heat units does not require the use of a constant because all x-rays possess the same ionizing potential d. 1.41, 1.0, 1.35 | back 71 D. 1.41, 1.0, 1.35 -Fill-in questions may be used on the certification exam, although they probably will not have three blanks like this one. Although we do not often deal with heat units, be sure to review the constants used in heat unit production. |
front 72 When a quality control test is performed to ensure that the penetrating ability of the x-ray beam is accurate, the result must be within what amount of the control panel setting? a. 2% of SID b. 4% c. 10% d. 50% | back 72 D.50% -kVp must be within plus or minus 10% of what is set. Be aware that some sources set the range from 4-6 kVp. |
front 73 The primary type of grid used in diagnostic imaging is: a. Crosshatch b. Parallel c. Rhombic d. Focused | back 73 D. Focused |
front 74 The portion of the image-intensifier tube that converts electron energy to visible light is the: a. Output phosphor b. Photocathode c. Input phosphor d. Detector element | back 74 A. Focused -This is where the electronic image again becomes visible. |
front 75 The portion of the image-intensifier tube that converts visible light to electrons is the: a. Output phosphor b. Photocathode c. Input phosphor d. Anode | back 75 B. Photocathode -The electrons are then accelerated toward the output phosphor. |
front 76 The input phosphor of the image-intensifier tube converts: a. Electron energy to x-ray energy b. X-rays and heat to visible light c. X-ray energy to visible light d. X-ray energy to an electronic image | back 76 C. X-ray energy to visible light -This is the point where the rays exiting the patient strike the input phosphor and it glows with visible light, releasing electrons from the photocathode. |
front 77 Total brightness gain achieved using an image intensifier equals: a. Flux gain times minification gain b. Diameter of input phosphor times diameter of output phosphor c. Intensification factor—brightness without an image intensifier divided by brightness with an image intensifier d. Total light emitted at the photocathode | back 77 A. Flux gain times minification gain -This is the equation used to calculate total brightness gain. |
front 78 Single-phase, full-wave rectification produces: a. Direct current b. Pulsating direct current c. Pulsating direct current with 120 pulses per second d. Pulsating direct current with 120 pulses per second and 100% ripple | back 78 D. Pulsating direct current with 120 pulses per second and 100% ripple -Although A, B, and C are all good answers, remember that we need to choose the one best answer, which is D. You are not likely to encounter this type of wording, but this question is a good exercise to remind you to look for the one best answer. |
front 79 Three-phase, 6-pulse full-wave rectification produces: a. Direct current with 13% ripple b. Direct current with 4% ripple c. Direct current with 100% ripple d. Alternating current with 13% ripple | back 79 A. Direct current with 13% ripple -13% ripple direct current, not alternating current. Three-phase, 12-pulse produces ripple as low as 4%. Single-phase ripple is 100% |
front 80 Three-phase, 12-pulse full-wave rectification produces: a. Direct current with 13% ripple b. Direct current with 4% ripple c. Direct current with 100% ripple d. Alternating current with 13% ripple | back 80 B. Direct current with 4% ripple -Three-phase, six-pulse produces ripple as low as 13%. Single-phase ripple is 100%. |
front 81 The increase in average photon energy when using three-phase, 6-pulse equipment compared with single phase equipment is: a. 1.35% b. 41% c. 1.41% d. 35% | back 81 D. 35% -This is why the formula for heat units for three phase, six-pulse equipment is multiplied by 1.35. |
front 82 The increase in average photon energy when using three phase, 12-pulse equipment compared with single-phase equipment is: a. 1.35% b. 41% c. 1.41% d. 35% | back 82 B. 41% -This is why the formula for calculating heat units for three-phase, 12-pulse equipment is multiplied by 1.41. |
front 83 Programs that deal with the safe and reliable operation of equipment and programs that address all aspects of the delivery of radiology services are called, respectively: a. Quality assurance and quality control b. Total quality improvement c. Quality control and quality assurance d. Total quality management | back 83 C. Quality control and quality assurance -This question asks for two answers; be sure you know what is meant by each term. |
front 84 An image intensifier’s conversion factor is expressed as the ratio of: a. The flux gain to the minification gain b. The luminance of the output phosphor to the exposure rate at the input phosphor c. The exposure rate at the output phosphor to the exposure rate at the input phosphor d. The size of the input phosphor to the size of the output phosphor | back 84 B. The luminance of the output phosphor to the exposure rate at the input phosphor |
front 85 The collimator must be accurate to a level of: a. ±4% of SID b. ±5% of SID c. ±2% of SID d. ±10% of SID | back 85 C. ±2% of SID -Be sure to review these quality control standards before you take the certification exam. |
front 86 kVp must be accurate to within: a. 1 of the amount on the control panel setting b. 5% of the control panel setting c. 2% of SID d. 10% of the control panel setting | back 86 D. 10% of the control panel setting -Some sources state accuracy within 4-6 kVp |
front 87 Exposure linearity must be accurate to within: a. 4 b. 5% c. 2% of SID d. 10% | back 87 B. 10% |
front 88 Exposure reproducibility must be accurate to within: a. 4 b. 5% c. 2% of SID d. 10% | back 88 B. 5% |
front 89 Fluoroscopy exposure rate may not exceed: a. 10 mGya per minute b. 100 mGya per minute c. 5 mGya per minute d. 1 mGya minute | back 89 B. 100 mGya per minute -It is measured in-air at the tabletop. |
front 90 The feature that automatically adjusts kVp and mA during fluoroscopy is called: a. AEC b. ABC c. DQE d. SNR | back 90 B. ABC -Automatic brightness control |
front 91 The test that measures the accuracy of adjacent mA stations is: a. Exposure reproducibility b. Spinning top test c. Pinhole camera d. Exposure linearity | back 91 D. Exposure linearity |
front 92 The test that measures the accuracy of successive exposures is: a. Exposure reproducibility b. Spinning top test c. Pinhole camera d . Exposure linearity | back 92 A. Exposure reproducibility |
front 93 The accuracy of the timer must be within: a. 10% of the time chosen for exposures over 10 ms b. 5% of the time chosen for exposures over 10 ms c. 5% of the time chosen for exposures under 10 ms d. 10% of the time chosen for exposures under 10 ms | back 93 B. 5% of the time chosen for exposures over 10 ms |
front 94 Resolution of the television system may be measured using the following tool(s): a. Wire mesh test b. Line pairs/millimeter resolution tool along with a kVp meter c. Resolution test pattern d. Resolution test pattern, wire mesh test, or line pairs/millimeter resolution tool | back 94 C. Resolution test pattern |
front 95 AECs may be tested using: a. Phantoms b. Images of real patients c. Analog meters d. Fluoroscopic screens | back 95 A. Phantoms -This provides for actually radiographing an object |
front 96 The number of mAs used for analog fluoroscopy is: a. 300 to 500 b. 3 to 5 c. 10 to 12 d. 100 to 300 | back 96 B. 3 to 5 -Note that this question is asking the amount of mA actually used and not the setting of the mA on the control panel. |
front 97 Marks on the focal track of the anode resulting from bombardment of electrons are called: a. Melts b. Bullet marks c. Pitting d. Cracks | back 97 C. Pitting -Choices A, B, and D are also marks on the focal track, but they are caused by malfunctions. |
front 98 Effective quality control and quality assurance programs are required for accreditation by: a. The Joint Commission b. Joint Review Committee on Education in Radiologic Technology c. American Healthcare Radiology Administrators d. American Registry of Radiologic Technologists | back 98 A. The Joint Commission -Representatives of The Joint Commission (TJC) regularly visit health care facilities to examine documentation of compliance with quality control and quality assurance standards. |
front 99 What regulates the duration of x-ray production? a. Step-down transformer b. Rectifier c. Cathode d. Timer | back 99 D. Timer -Most equipment uses an electronic timer. |
front 100 Choose all that are properties of x-rays (choose 6): a. Electrically negative b. Affect film emulsion c. Scatter and produce secondary radiation d. Invisible to the human eye e. Travel at the speed of light (186,000 miles per hour) f. Possess wavelengths between 1 Å and 5 Å g. Travel in bundles of energy called photons h. Can ionize matter and gases i. Can be focused by collimators j. Cause phosphors to fluoresce | back 100 B. Affect film emulsion C. Scatter and produce secondary radiation D. Invisible to the human eye G. Travel in bundles of energy called photons H.Can ionize matter and gases J.Cause phosphors to fluoresce |