front 1 The process of "leveling and windowing" of digital images determines the image
| back 1 B.
|
front 2 The relationship between the intensity of light striking a film and the intensity of light transmitted through the film is an expression of which of the following?
| back 2 B.
|
front 3 As the CR laser scanner/reader recognizes the phosphostimulated luminescence (PSL) released by the PSP storage plate, it constructs a graphic representation of pixel value distribution called a
| back 3 B.
|
front 4 The function(s) of the developer in film processing is (are) to
| back 4 B.
|
front 5 Cassetteless digital systems have a fixed spatial resolution determined by:
| back 5 C.
|
front 6 Which of the following statements regarding dual x-ray absorptiometry is (are) true?
| back 6 C.
|
front 7 Which section of the automatic processor shown in Figure 6–9 is associated with preservation of the x-ray image?
| back 7 C.
|
front 8 Due to the high sensitivity of digital detectors to low intensity radiation (background, scatter and/or off-focus radiation), there is likely to be scatter and off-focus radiation contributing to the image outside the collimation margins. Since many radiologists find this distracting, the most appropriate radiographer action would be to:
| back 8 B.
|
front 9 The luminescent light emitted by the PSP is transformed into the image seen on the CRT by the
| back 9 C.
|
front 10 Which of the two film emulsions illustrated in Figure 4–11 requires more exposure to produce at density of 1.75?
| back 10 B.
|
front 11 A change from 400-speed screens to 200-speed screens would require what change in milliampere-seconds?
| back 11 C.
|
front 12 Histogram data that is skewed relative to the values of interest (VOI) of the histogram analysis used for a particular exam may be caused by all of the following, except:
| back 12 B.
|
front 13 Which of the following affects both radiographic density and intensifying screen speed?
| back 13 B.
|
front 14 An exposed image plate will retain its original image quality for about
| back 14 B.
|
front 15 The radiograph in Figure 6–12 could be improved in which of the following ways?
| back 15 C.
|
front 16 The pre-contrast image used to subtract pixel values from the post-contrast image showing contrast-filled blood vessels in digital subtraction angiography is called the:
| back 16 A.
|
front 17 What effect will a stained intensifying screen have on the finished radiograph?
| back 17 C.
|
front 18 The presence of dust or scratches on intensifying screens will cause
| back 18 A.
|
front 19 Unopened boxes of radiographic film should be stored away from radiation and
| back 19 B.
|
front 20 An important feature of the pixel in a flat-panel TFT digital detector active matrix array is the:
| back 20 B.
|
front 21 Which of the following is (are) tested as part of a quality assurance (QA) program?
| back 21 D.
|
front 22 Which of the following is (are) methods used for x-ray film silver reclamation?
| back 22 C.
|
front 23 Factors that contribute to film fog include
| back 23 D.
|
front 24 Silver reclamation may be accomplished in which of the following ways?
| back 24 D.
|
front 25 In which section of the automatic processor shown in Figure 7–9 are the exposed silver halide crystals changed to black metallic silver?
| back 25 A.
|
front 26 Which of the following possesses the widest dynamic range?
| back 26 D.
|
front 27 Which of the following requires two exposures to evaluate focal-spot accuracy?
| back 27 B.
|
front 28 Figure 4–24 is representative of
| back 28 C.
|
front 29 Exposure-type artifacts include
| back 29 B.
|
front 30 The term windowing describes the practice of
| back 30 B.
|
front 31 All of the following are potential digital pre-processing problems, except:
| back 31 A.
|
front 32 Which of the following contribute(s) to base-plus fog?
| back 32 D.
|
front 33 While indirect digital detectors use a scintillator (phosphor) to convert X-ray energy, direct detectors use a:
| back 33 A.
|
front 34 A wire-mesh test is performed to evaluate screen
| back 34 B.
|
front 35 The term spectral matching refers to the fact that film sensitivity must be matched with the
| back 35 A.
|
front 36 Which of the following materials may be used as grid interspace material?
| back 36 C.
|
front 37 What information, located on each box of film, is important to note and has a direct relationship to image quality?
| back 37 C.
|
front 38 Boxes of film stored in too warm an area may be subject to
| back 38 B.
|
front 39 The function of the developer solution chemicals is to
| back 39 C.
|
front 40 When a slow screen–film system is used with a fast screen–film AEC system, the resulting images
| back 40 A.
|
front 41 Geometric blur can be evaluated using all the following devices except
| back 41 C.
|
front 42 Which of the two sensitometric curves shown in Figure 7–8 requires more exposure to produce a density of 2.0 on the finished radiograph?
| back 42 B.
|
front 43 The X-ray scintillator layer used in in direct flat-panel digital
detectors is usually either _____________ or ______________. | back 43 B. |
front 44 The cause of films coming from the automatic processor still damp can be
| back 44 C.
|
front 45 The functions of a picture archiving and communication system (PACS) include
| back 45 C.
|
front 46 The changes between the images shown in Figure 4–7 represent changes made to
| back 46 C.
|
front 47 Incorrect spectral matching between rare earth intensifying screens and film emulsion results in
| back 47 B.
|
front 48 During measurement of blood pressure, which of the following occurs as the radiographer controls arterial tension with the sphygmomanometer?
| back 48 B.
|
front 49 The purpose of the automatic processor's transport system is to
| back 49 C.
|
front 50 Which of the lines indicated in Figure 7–1 correctly demonstrates the relationship between the exposure received by the PSP and its resulting luminescence as it is laser scanned?
| back 50 A.
|
front 51 "Saturation" of an image in CR means that:
| back 51 B.
|
front 52 Which of the following can cause poor screen–film contact?
| back 52 D.
|
front 53 X-ray film emulsion is most sensitive to safelight fog
| back 53 B.
|
front 54 Off-focus and scatter radiation outside of the exposure field when using CR can cause:
| back 54 B.
|
front 55 Inadequate collimation in CR imaging can result in an image that is too
| back 55 D.
|
front 56 Exposed silver halide crystals are changed to black metallic silver by the
| back 56 B.
|
front 57 Which of the following can affect the amount of developer replenisher delivered per film in an automatic processor?
| back 57 C.
|
front 58 Before a flat-panel detector can be used for a radiographic exposure, it must be prepared. This preparation is referred to as:
| back 58 B.
|
front 59 The device used to give film emulsion a predetermined exposure in order to test its response to processing is called the
| back 59 A.
|
front 60 A fill factor of 80% in direct or indirect digital radiography means that:
| back 60 A.
|
front 61 A film emulsion having wide latitude is likely to exhibit
| back 61 D.
|
front 62 The chest radiograph shown in Figure 4–1 demonstrates
| back 62 C.
|
front 63 The optimal alignment of the anatomical part being imaged for all digital receptors should be:
| back 63 D.
|
front 64 The x-ray tube used in CT must be capable of
| back 64 D.
|
front 65 Sampling frequency in computed radiography (CR) is expressed as:
| back 65 D.
|
front 66 An unexposed and processed film will have a density of about
| back 66 B.
|
front 67 With a given exposure, as intensifying screen speed increases, how is radiographic density affected?
| back 67 B.
|
front 68 Which of the following is the correct order of radiographic film processing?
| back 68 C.
|
front 69 If obtaining multiple images on one image plate, it is important to:
| back 69 C.
|
front 70 Base-plus fog is a result of
| back 70 D.
|
front 71 Component part(s) of x-ray film include which of the following?
| back 71 C.
|
front 72 Resolution in CR increases as
| back 72 C.
|
front 73 Pixel size and spacing determine the spatial resolution of the digital image. This is known as:
| back 73 A.
|
front 74 Which of the following digital post-processing methods remove high-frequency noise from the image?
| back 74 C.
|
front 75 The safe approach to avoid an exposure field recognition error when using CR is to:
| back 75 A.
|
front 76 A film emerging from the automatic processor exhibits excessive density. This might be attributable to the following
| back 76 B.
|
front 77 Film base is currently made of which of the following materials?
| back 77 C.
|
front 78 The device shown in Figure 4–16 is used for
| back 78 C.
|
front 79 Which of the following is an abnormal intensifying screen action?
| back 79 D.
|
front 80 What information must be included on an x-ray image for it to be considered as legitimate legal evidence?
| back 80 B.
|
front 81 Characteristics of DR imaging include
| back 81 D.
|
front 82 Chemical fog may be attributed to
| back 82 D.
|
front 83 The artifacts on the radiograph shown in Figure 4–25 are called
| back 83 B.
|
front 84 In digital imaging, the maximum spatial resolution is equal to:
| back 84 A.
|
front 85 For the same FOV, spatial resolution will be improved using
| back 85 B.
|
front 86 The processor rollers that are out of solution and function to transfer the film from one solution to another are the
| back 86 C.
|
front 87 In which section of the automatic processor seen in the figure below are the unexposed silver halide crystals removed from the emulsion?
| back 87 B.
|
front 88 When radiographing a cross-table lateral hip or axial shoulder using CR, one method of creating a collimation margin at the bottom of the radiograph is to:
| back 88 A.
|
front 89 The automatic film processor's recirculation system functions to
| back 89 C.
|
front 90 The instrument that is used frequently in quality-control programs to measure varying degrees of x-ray exposure is the
| back 90 C.
|
front 91 What determines the quantity of fluorescent light emitted from a fluorescent screen?
| back 91 D.
|
front 92 What is the purpose of the thin layer of lead that is often located behind the rear intensifying screen in an IR?
| back 92 D.
|
front 93 All of the following are factors that cause low contrast in CR images, except:
| back 93 C.
|
front 94 The smallest digital detectors (approximately 100 microns) provide the best spatial resolution and, therefore, are best-suited for use in:
| back 94 B.
|
front 95 IRs/cassettes frequently have a lead-foil layer behind the rear screen that functions to
| back 95 B.
|
front 96 Objectionable widening of the histogram in CR can be caused by all of the following, except:
| back 96 D.
|
front 97 One reason why only one image is preferred per image plate is:
| back 97 B.
|
front 98 A QA program serves to
| back 98 D.
|
front 99 As the conversion efficiency of intensifying-screen phosphors increases,
| back 99 C.
|
front 100 Which apparatus is needed for the construction of a sensitometric curve?
| back 100 A.
|
front 101 In order to avoid background radiation artifacts when using CR, it is important to:
| back 101 F.
|
front 102 Phosphors suitable for use in intensifying screens should have which of the following characteristics?
| back 102 B.
|