Print Options

Card layout:

← Back to notecard set|Easy Notecards home page

Instructions for Side by Side Printing
  1. Print the notecards
  2. Fold each page in half along the solid vertical line
  3. Cut out the notecards by cutting along each horizontal dotted line
  4. Optional: Glue, tape or staple the ends of each notecard together
  1. Verify Front of pages is selected for Viewing and print the front of the notecards
  2. Select Back of pages for Viewing and print the back of the notecards
    NOTE: Since the back of the pages are printed in reverse order (last page is printed first), keep the pages in the same order as they were after Step 1. Also, be sure to feed the pages in the same direction as you did in Step 1.
  3. Cut out the notecards by cutting along each horizontal and vertical dotted line
Print these notecards...Print as a list

282 notecards = 71 pages (4 cards per page)

Viewing:

Radreview technical factors 281

front 1

Focal-spot blur is greatest
A.
directly along the course of the central ray
B.
toward the cathode end of the x-ray beam
C.
toward the anode end of the x-ray beam
D.
as the SID is increased

back 1

B.
toward the cathode end of the x-ray beam

front 2

Accurate operation of the AEC device depends on

the thickness and density of the object

positioning of the object with respect to the photocell

beam restriction

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 2

D.
1, 2, and 3

front 3

An x-ray exposure of a particular part is made and restricted to a 14 x 17 in. field size. The same exposure is repeated, but the x-ray beam is restricted to a 4 x 4 in. field. Compared with the first image, the second image will demonstrate

less density

more contrast

more density

A.
1 only
B.
1 and 2 only
C.
3 only
D.
2 and 3 only

back 3

B.
1 and 2 only

front 4

If a particular grid has lead strips 0.40 mm thick, 4.0 mm high, and 0.25 mm apart, what is its grid ratio?
A.
8:1
B.
10:1
C.
12:1
D.
16:1

back 4

D.
16:1

front 5

An exposure was made at 40-in. SID using 5 mAs and 105 kVp with an 8:1 grid. In an effort to improve radiographic contrast, the image is repeated using a 12:1 grid and 90 kVp. Which of the following exposure times will be most appropriate, using 400 mA, to maintain the original density?
A.
0.01 s
B.
0.03 s
C.
0.1 s
D.
0.3 s

back 5

B.
0.03 s

front 6

Compared with slow-speed screens, high-speed screens are often used to

minimize patient dose

keep exposure time to a minimum

image fine anatomic details

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 6

B.
1 and 2 only

front 7

Which of the following technical changes would best serve to remedy the effect of very dissimilar tissue densities?
A.
Use of high-speed screens
B.
Use of a high-ratio grid
C.
High-kilovoltage exposure factors
D.
High milliampere-seconds exposure factors

back 7

C.
High-kilovoltage exposure factors

front 8

Tree-like branching black marks on a radiograph are usually the result of
A.
bending the film acutely
B.
improper development
C.
improper film storage
D.
static electricity

back 8

D.
static electricity

front 9

Recorded detail depends on all the following except

1. quantity of filtration.

2. anode angle.

3. intensification factor of screens.
A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 9

A.
1 only

front 10

The exposure factors of 300 mA, 0.017 second, and 72 kVp produce an mAs value of
A.
5.
B.
50.
C.
500.
D.
5000.

back 10

A.
5

front 11

In which of the following examinations would a cassette front with very low absorption properties be especially desirable?
A.
Extremity radiography
B.
Abdominal radiography
C.
Mammography
D.
Angiography

back 11

C.
Mammography

front 12

An exposure was made at a 36-in. SID using 300 mA, a 30-ms exposure, and 80 kV with a 400-speed film–screen combination and an 8:1 grid. It is desired to repeat the radiograph and, in order to improve recorded detail, to use a 40-in. SID and a 200-speed film–screen combination. With all other factors remaining constant, what exposure time will be required to maintain the original radiographic density?
A.
0.03 second
B.
0.07 second
C.
0.14 second
D.
0.36 second

back 12

B.
0.07 second

front 13

Involuntary motion can be caused by
1. peristalsis.
2. severe pain.
3. heart muscle contraction.

A.
1 only
B.
2 only
C.
1 and 2 only
D.
1, 2, and 3

back 13

D.
1, 2, and 3

front 14

A satisfactory radiograph was made using a 36-in. SID, 12 mAs, and a 12:1 grid. If the examination will be repeated at a distance of 42 in. and using a 5:1 grid, what should be the new milliampere-seconds value to maintain the original density?
A.
5.6
B.
6.5
C.
9.7
D.
13

back 14

B.
6.5

front 15

A satisfactory radiograph was made using a 40-inch SID, 10 mAs, and a 12:1 grid. If the examination will be repeated at a distance of 48 inches and using an 8:1 grid, what should be the new mAs to maintain the original density?
A.
5.6
B.
8.8
C.
11.5
D.
14.4

back 15

C.
11.5

front 16

If a 6-in. OID is introduced during a particular radiographic examination, what change in SID will be necessary to overcome objectionable magnification?
A.
The SID must be increased by 6 in..
B.
The SID must be increased by 18 in..
C.
The SID must be decreased by 6 in..
D.
The SID must be increased by 42 in..

back 16

D.
The SID must be increased by 42 in..

front 17

Angulation of the central ray may be required
1. to avoid superimposition of overlying structures.
2. to avoid foreshortening or self-superimposition.
3. to project through certain articulations.

A.
1 only
B.
2 only
C.
1 and 3 only
D.
1, 2, and 3

back 17

D.
1, 2, and 3

front 18

Types of moving grid mechanisms include
1. oscillating.
2. reciprocating.
3. synchronous.

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
2 and 3 only

back 18

B.
1 and 2 only

front 19

What is the best way to reduce magnification distortion?
A.
Use a small focal spot.
B.
Increase the SID.
C.
Decrease the OID.
D.
Use a slow screen-film combination.

back 19

C.
Decrease the OID.

front 20

The factors that control recorded detail include
1. Focal spot size
2. Type of rectification
3. SID

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 20

B.
1 and 3 only

front 21

Which of the following groups of exposure factors will produce the shortest scale of contrast?
A.
200 mA, 0.08 second, 95 kV, 12:1 grid
B.
500 mA, 0.03 second, 70 kV, 8:1 grid
C.
300 mA, 0.05 second, 95 kV, 8:1 grid
D.
600 mA, 1/40 seconds, 70 kV, 6:1 grid

back 21

B.
500 mA, 0.03 second, 70 kV, 8:1 grid

front 22

Characteristics of high-ratio focused grids, compared with lower-ratio grids, include which of the following?

They allow more positioning latitude.

They are more efficient in collecting SR.

They absorb more of the useful beam.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 22

C.
2 and 3 only

front 23

A part whose width is 6 inches will be imaged at 44 inches SID. The part to be imaged lies 9 inches from the IR. What will be the projected image width of the part?
A.
8 inches
B.
10 inches
C.
12 inches
D.
20 inches

back 23

A.
8 inches

front 24

Geometric unsharpness is influenced by which of the following?

Distance from object to image

Distance from source to object

Distance from source to image

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 24

D.
1, 2, and 3

front 25

Which of the following will influence recorded detail?

Screen speed

Screen–film contact

Focal spot

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 25

D.
1, 2, and 3

front 26

All the following are related to recorded detail except
A.
milliamperage
B.
focal-spot size
C.
screen speed
D.
OID

back 26

A.
milliamperage

front 27

The exposure factors of 400 mA, 70 ms, and 78 kV were used to produce a particular radiographic density and contrast. A similar radiograph can be produced using 500 mA, 90 kV, and
A.
14 ms
B.
28 ms
C.
56 ms
D.
70 ms

back 27

B.
28 ms

front 28

Which of the following terms is used to express resolution/recorded detail?
A.
Kiloelectronvolts (keV)
B.
Modulation transfer function (MTF)
C.
Relative speed
D.
Latitude

back 28

B.
Modulation transfer function (MTF)

front 29

A radiograph exposed using a 12:1 ratio grid may exhibit a loss of density at its lateral edges because the
A.
SID was too great.
B.
grid failed to move during the exposure.
C.
x-ray tube was angled in the direction of the lead strips.
D.
CR was off-center.

back 29

A.
SID was too great.

front 30

Characteristics of low ratio focused grids include the following:
1. they have a greater focal range
2. they are less efficient in collecting SR
3. they can be used inverted

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 30

B.
1 and 2 only

front 31

Which of the following terms refers to light being reflected from one intensifying screen, through the film, to the opposite emulsion and intensifying screen?
A.
Reflectance
B.
Crossover
C.
Scatter
D.
Filtration

back 31

B.
Crossover

front 32

Shape distortion is influenced by the relationship between the

x-ray tube and the part to be imaged

part to be imaged and the IR

IR and the x-ray tube

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 32

D.
1, 2, and 3

front 33

The sensitometric curve may be used to

identify automatic processing problems

determine film sensitivity

illustrate screen speed

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 33

B.
1 and 2 only

front 34

Which of the following units is (are) used to express resolution?

Line-spread function

Line pairs per millimeter

Line-focus principle

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 34

B.
1 and 2 only

front 35

The fact that x-ray intensity across the primary beam can vary as much as 45% describes the
A.
line-focus principle.
B.
transformer law.
C.
anode heel effect.
D.
inverse-square law.

back 35

C.
anode heel effect.

front 36

An exposure was made using 300 mA, 40 ms exposure, and 85 kV. Each of the following changes will decrease the radiographic density by one half except a change to
A.
1/50 s exposure
B.
72 kV
C.
10 mAs
D.
150 mA

back 36

C.
10 mAs

front 37

Which of the following is the factor of choice for the regulation of radiographic (optical) density?
A.
kVp
B.
mAs
C.
SID
D.
Filtration

back 37

B.
mAs

front 38

Exposure factors of 90 kVp and 4 mAs are used for a particular nongrid exposure. What should be the new mAs if an 8:1 grid is added?
A.
8
B.
12
C.
16
D.
20

back 38

C.
16

front 39

The interaction between x-ray photons and matter illustrated in Figure 4–22 is most likely to be associated with

high kilovoltage

high contrast

high-atomic-number absorber

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3
Figure 4–22.

back 39

A.
1 only

front 40

A decrease in kilovoltage will result in
A.
a decrease in optical density
B.
a decrease in contrast
C.
a decrease in recorded detail
D.
a decrease in image resolution

back 40

A.
a decrease in optical density

front 41

Using a short (25–30 in.) SID with a large (14 x 17 in.) IR is likely to
A.
increase the scale of contrast
B.
increase the anode heel effect
C.
cause malfunction of the AEC
D.
cause premature termination of the exposure

back 41

B.
increase the anode heel effect

front 42

When blue-emitting rare earth screens are matched properly with the correct film emulsion, what type of safelight should be used in the darkroom?

Red

GBX-2

Ultraviolet

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 42

B.
1 and 2 only

front 43

The reduction in x-ray photon intensity as the photon passes through material is termed
A.
absorption
B.
scattering
C.
attenuation
D.
divergence

back 43

C.
attenuation

front 44

A radiograph made using 300 mA, 0.1 second, and 75 kVp exhibits motion unsharpness, but otherwise satisfactory technical quality. The radiograph will be repeated using a shorter exposure time. Using 86 kV and 500 mA, what should be the new exposure time?
A.
0.12 second
B.
0.06 second
C.
0.03 second
D.
0.01 second

back 44

C.
0.03 second

front 45

With milliamperage adjusted to produce equal exposures, all the following statements are true except
A.
a single-phase examination done at 10 mAs can be duplicated with three-phase, 12-pulse at 5 mAs.
B.
There is greater patient dose with three-phase equipment than with single-phase equipment.
C.
Three-phase equipment can produce comparable radiographs with less heat unit (HU) buildup.
D.
Three-phase equipment produces lower-contrast radiographs than single-phase equipment.

back 45

B.
There is greater patient dose with three-phase equipment than with single-phase equipment.

front 46

Which of the following terms is used to describe unsharp edges of tiny radiographic details?
A.
Diffusion
B.
Mottle
C.
Blur
D.
Umbra

back 46

C.
Blur

front 47

Focusing distance is associated with which of the following?
A.
Computed tomography
B.
Chest radiography
C.
Magnification radiography
D.
Grids

back 47

D.
Grids

front 48

Phosphors classified as rare earth include

lanthanum oxybromide.

gadolinium oxysulfide.

cesium iodide.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 48

B.
1 and 2 only

front 49

Greater latitude is available to the radiographer when using

high-kilovoltage factors.

a slow film–screen combination.

a high-ratio grid.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 49

B.
1 and 2 only

front 50

The relationship between the height of a grid's lead strips and the distance between them is referred to as grid
A.
ratio
B.
radius
C.
frequency
D.
focusing distance

back 50

A.
ratio

front 51

Exposure factors of 400 mA, 20 ms, 68 kVp, 400-speed system, at 40-in. SID were used to produce a satisfactory radiographic image. A change to 4 mAs can be best compensated for by which of the following?
A.
Increasing the SID to 60 in.
B.
Decreasing the SID to 20 in.
C.
Decreasing the speed to 200
D.
Increasing the kilovoltage to 78 kVp

back 51

D.
Increasing the kilovoltage to 78 kVp

front 52

Exposure rate increases with an increase in
1. mA.
2. kVp.
3. SID.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 52

B.
1 and 2 only

front 53

Of the following groups of technical factors, which will produce the greatest radiographic density?
A.
10 mAs, 74 kV, 44-in. SID
B.
10 mAs, 74 kV, 36-in. SID
C.
5 mAs, 85 kV, 48-in. SID
D.
5 mAs, 85 kV, 40-in. SID

back 53

B.
10 mAs, 74 kV, 36-in. SID

front 54

Radiographic contrast is the result of
A.
transmitted electrons
B.
differential absorption
C.
absorbed photons
D.
milliampere-seconds selection

back 54

B.
differential absorption

front 55

X-ray photon energy is inversely related to

photon wavelength

applied milliamperes (mA)

applied kilovoltage (kV)

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 55

A.
1 only

front 56

A lateral radiograph of the lumbar spine was made using 200 mA, 1-second exposure, and 90 kV. If the exposure factors were changed to 200 mA, 0.5 second, and 104 kV, there would be an obvious change in which of the following?

Radiographic density

Scale of radiographic contrast

Distortion

A.
1 only
B.
2 only
C.
2 and 3 only
D.
1, 2, and 3

back 56

B.
2 only

front 57

If the center photocell were selected for a lateral projection of the lumbar spine that was positioned with the spinous processes instead of the vertebral bodies centered to the grid, how would the resulting radiograph look?
A.
The image would be underexposed.
B.
The image would be overexposed.
C.
The image would be correctly exposed.
D.
An exposure could not be made.

back 57

A.
The image would be underexposed.

front 58

Conditions contributing to poor radiographic film archival quality include

fixer retention

insufficient developer replenishment

poor storage conditions

A.
1 only
B.
3 only
C.
2 and 3 only
D.
1, 2, and 3

back 58

D.
1, 2, and 3

front 59

Geometric unsharpness is most likely to be greater
A.
at long SIDs.
B.
at the anode end of the image.
C.
with small focal spots.
D.
at the cathode end of the image.

back 59

D.
at the cathode end of the image.

front 60

A 5-in. object to be radiographed at a 44-in. SID lies 6 in. from the IR. What will be the image width?
A.
5.1 in.
B.
5.7 in.
C.
6.1 in.
D.
6.7 in.

back 60

B.
5.7 in.

front 61

Disadvantages of using low-kilovoltage technical factors include

insufficient penetration

increased patient dose

diminished latitude

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 61

D.
1, 2, and 3

front 62

What pixel size has a 1024 x 1024 matrix with a 35-cm FOV?
A.
30 mm
B.
0.35 mm
C.
0.15 mm
D.
0.03 mm

back 62

B.
0.35 mm

front 63

Which combination of exposure factors most likely will contribute to producing the shortest-scale contrast?
A.
mAs: 10; kV: 70; Film–screen system: 400; Grid ratio: 5:1; Field size: 14 x 17 in.
B.
mAs: 12; kV: 90; Film–screen system: 200; Grid ratio: 8:1; Field size: 14 x 17 in.
C.
mAs: 15; kV: 90; Film–screen system: 200; Grid ratio: 12:1; Field size: 11 x 14 in.
D.
mAs: 20; kV: 80; Film–screen system: 400; Grid ratio: 10:1; Field size: 8 x 10 in.

back 63

D.
mAs: 20; kV: 80; Film–screen system: 400; Grid ratio: 10:1; Field size: 8 x 10 in.

front 64

Which of the following will produce the greatest distortion?
A.
AP projection of the skull
B.
PA projection of the skull
C.
37° AP axial of the skull
D.
20° PA axial of the skull

back 64

C.
37° AP axial of the skull

front 65

Terms that refer to size distortion include

magnification

attenuation

elongation

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 65

A.
1 only

front 66

The exposure factors used for a particular nongrid x-ray image were 300 mA, 4 ms, and 90 kV. Another image, using an 8:1 grid, is requested. Which of the following groups of factors is most appropriate?
A.
400 mA, 3 ms, 110 kV
B.
400 mA, 12 ms, 90 kV
C.
300 mA, 8 ms, 100 kV
D.
200 mA, 240 ms, 90 kV

back 66

B.
400 mA, 12 ms, 90 kV

front 67

Figure 4–19 is representative of

Figure 4–19.

A.
the anode heel effect
B.
the line-focus principle
C.
the inverse-square law
D.
the reciprocity law

back 67

A.
the anode heel effect

front 68

Which of the following is (are) characteristic(s) of a 16:1 grid?

It absorbs more useful radiation than an 8:1 grid.

It has more centering latitude than an 8:1 grid.

It is used with higher-kilovoltage exposures than an 8:1 grid.

A.
1 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 68

B.
1 and 3 only

front 69

In which of the following ways can higher radiographic contrast be obtained in abdominal radiography?
1. By using lower kilovoltage
2. By using a contrast medium
3. By limiting the field size

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 69

D.
1, 2, and 3

front 70

The line-focus principle expresses the relationship between
A.
the actual and the effective focal spot
B.
exposure given the IR and resulting density
C.
SID used and resulting density
D.
grid ratio and lines per inch

back 70

A.
the actual and the effective focal spot

front 71

An increase from 78 to 92 kVp will result in an increase in which of the following?

Latitude

Scale of grays

Optical density

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 71

D.
1, 2, and 3

front 72

Using fixed milliampere-seconds and variable kilovoltage technical factors, each centimeter increase in patient thickness requires what adjustment in kilovoltage?
A.
Increase 2 kV
B.
Decrease 2 kV
C.
Increase 4 kV
D.
Decrease 4 kV

back 72

A.
Increase 2 kV

front 73

An exposure was made using 600 mA and 18 ms. If the mA is changed to 400, which of the following exposure times would most closely approximate the original radiographic density?
A.
16 ms
B.
0.16 second
C.
27 ms
D.
0.27 second

back 73

C.
27 ms

front 74

What pixel size has a 2048 x 2048 matrix with an 80-cm FOV?
A.
0.04 mm
B.
0.08 mm
C.
0.2 mm
D.
0.4 mm

back 74

D.
0.4 mm

front 75

In electronic imaging, as digital image matrix size increases

pixel size decreases

resolution decreases

pixel depth decreases

A.
1 only
B.
2 only
C.
1 and 2 only
D.
2 and 3 only

back 75

A.
1 only

front 76

Which of the following would be appropriate cassette front material(s)?

Tungsten

Magnesium

Bakelite

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 76

C.
2 and 3 only

front 77

OID is related to recorded detail in which of the following ways?
A.
Radiographic detail is directly related to OID.
B.
Radiographic detail is inversely related to OID.
C.
As OID increases, so does radiographic detail.
D.
OID is unrelated to radiographic detail.

back 77

B.
Radiographic detail is inversely related to OID.

front 78

In an AP abdomen radiograph taken at 105-cm SID during an IVU series, one renal shadow measures 9 cm in width. If the OID is 18 cm, what is the actual width of the kidney?
A.
5 cm
B.
7.5 cm
C.
11 cm
D.
18 cm

back 78

B.
7.5 cm

front 79

If 32 mAs and 50-speed screens were used to produce a particular radiographic density, what new mAs value would be required to produce the same density if the screen speed were changed to 400?
A.
4 mAs
B.
40 mAs
C.
175 mAs
D.
256 mAs

back 79

A.
4 mAs

front 80

Which of the following pathologic conditions would require an increase in exposure factors?
A.
Pneumoperitoneum
B.
Obstructed bowel
C.
Renal colic
D.
Ascites

back 80

D.
Ascites

front 81

Which of the following is (are) associated with subject contrast?

Patient thickness

Tissue density

Kilovoltage

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 81

D.
1, 2, and 3

front 82

If 300 mA has been selected for a particular exposure, what exposure time would be required to produce 60 mAs?
A.
1/60 second
B.
1/30 second
C.
1/10 second
D.
1/5 second

back 82

D.
1/5 second

front 83

According to the line-focus principle, an anode with a small angle provides

improved recorded detail.

improved heat capacity.

less heel effect.

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 83

A.
1 and 2 only

front 84

Low-kilovoltage exposure factors usually are indicated for radiographic examinations using

water-soluble, iodinated media

a negative contrast agent

barium sulfate

A.
1 only
B.
1 and 2 only
C.
3 only
D.
1 and 3 only

back 84

B.
1 and 2 only

front 85

What grid ratio is represented in Figure 4–8?

Figure 4–8.

A.
3:1
B.
5:1
C.
10:1
D.
16:1

back 85

C.
10:1

front 86

A part whose width is 6 inches will be imaged at 44 inches SID. The part to be imaged lies 9 inches from the IR. What will be the magnification factor?
A.
1.25
B.
1.86
C.
4.9
D.
7.3

back 86

A.
1.25

front 87

Which of the following is (are) characteristic(s) of a 5:1 grid?

It allows some positioning latitude.

It is used with high-kilovoltage exposures.

It absorbs a high percentage of scattered radiation.

A.
1 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 87

A.
1 only

front 88

The microswitch for controlling the amount of replenishment used in an automatic processor is located at the
A.
receiving bin
B.
crossover roller
C.
entrance roller
D.
replenishment pump

back 88

C.
entrance roller

front 89

A 15% increase in kVp accompanied by a 50% decrease in mAs will result in a(n)
A.
shorter scale of contrast.
B.
increase in exposure latitude.
C.
increase in radiographic density.
D.
decrease in recorded detail.

back 89

B.
increase in exposure latitude.

front 90

A dry laser printer is generally used when it is necessary to print digital images on film. This laser film is loaded under the following conditions
A.
daylight
B.
safelight with red filter
C.
safelight with amber filter
D.
total darkness

back 90

A.
daylight

front 91

Which of the following pathologic conditions would require an increase in exposure factors?
A.
Pneumoperitoneum
B.
Obstructed bowel
C.
Renal colic
D.
Ascites

back 91

D.
Ascites

front 92

What should be done to correct for magnification when using air-gap technique?
A.
Decrease OID
B.
Increase OID
C.
Decrease SID
D.
Increase SID

back 92

D.
Increase SID

front 93

In amorphous selenium flat-panel detectors, the term amorphous refers to a
A.
crystalline material having typical crystalline structure.
B.
crystalline material lacking typical crystalline structure.
C.
toxic crystalline material.
D.
homogeneous crystalline material.

back 93

B.
crystalline material lacking typical crystalline structure.

front 94

What pixel size has a 2,048 x 2,048 matrix with a 60-cm FOV?
A.
0.3 mm
B.
0.5 mm
C.
0.15 mm
D.
0.03 mm

back 94

A.
0.3 mm

front 95

An x-ray image of the ankle was made at 40-SID, 200 mA, 50 ms, 70 kV, 0.6 mm focal spot, and minimal OID. Which of the following modifications would result in the greatest increase in magnification?
A.
1.2 mm focal spot
B.
36-in. SID
C.
44-in. SID
D.
4-in. OID

back 95

D.
4-in. OID

front 96

Which of the following adult radiographic examinations usually require(s) use of a grid?

Ribs

Vertebrae

Shoulder

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 96

D.
1, 2, and 3

front 97

A particular radiograph was produced using 6 mAs and 110 kVp with an 8:1 ratio grid. The radiograph is to be repeated using a 16:1 ratio grid. What should be the new mAs?
A.
3
B.
6
C.
9
D.
12

back 97

C.
9

front 98

If 92 kV and 12 mAs were used for a particular abdominal exposure with single-phase equipment, what mAs would be required to produce a similar radiograph with three-phase, six-pulse equipment?
A.
36
B.
24
C.
8
D.
6

back 98

C.
8

front 99

Which of the following factors impact(s) recorded detail?

Focal-spot size

Subject motion

SOD

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 99

D.
1, 2, and 3

front 100

Recommended method(s) of minimizing motion unsharpness include

suspended respiration

short exposure time

patient instruction

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 100

D.
1, 2, and 3

front 101

Recommended method(s) of minimizing motion unsharpness include

suspended respiration

short exposure time

patient instruction

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 101

D.
1, 2, and 3

front 102

Geometric unsharpness will be least obvious

at long SIDs.

with small focal spots.

at the anode end of the image.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 102

D.
1, 2, and 3

front 103

If a radiograph were made of an average-size knee using automatic exposure control (AEC) and all three photocells were selected, the resulting radiograph would demonstrate
A.
excessive density.
B.
insufficient density.
C.
poor detail.
D.
adequate exposure.

back 103

B.
insufficient density.

front 104

Which of the following groups of exposure factors would be most appropriate to control involuntary motion?
A.
400 mA, 0.03 second
B.
200 mA, 0.06 second
C.
600 mA, 0.02 second
D.
100 mA, 0.12 second

back 104

C.
600 mA, 0.02 second

front 105

An exposure was made using 300 mA and 50 ms. If the exposure time is changed to 22 ms, which of the following milliamperage selections would most closely approximate the original radiographic density?
A.
300 mA
B.
400 mA
C.
600 mA
D.
700 mA

back 105

D.
700 mA

front 106

If a 4-inch collimated field is changed to a 14-inch collimated field, with no other changes, the radiographic image will possess
A.
more density/IR exposure.
B.
less density/IR exposure.
C.
more detail.
D.
less detail.

back 106

A.
more density/IR exposure.

front 107

Which of the following absorbers has the highest attenuation coefficient?
A.
Bone
B.
Muscle
C.
Fat
D.
Air

back 107

A.
Bone

front 108

An exposure was made using 600 mA, 0.04-s exposure, and 85 kVp. Each of the following changes will serve to decrease the radiographic density by one-half except change to
A.
1/50-s exposure
B.
72 kVp
C.
18 mAs
D.
300 mA

back 108

C.
18 mAs

front 109

Which of the following examinations might require the use of 120 kVp?
1. AP abdomen
2. Chest radiograph
3. Barium-filled stomach

A.
1 only
B.
2 only
C.
1 and 2 only
D.
2 and 3 only

back 109

D.
2 and 3 only

front 110

The radiograph of the pelvis shown in Figure 6–7 is unacceptable because of

Figure 6–7.

From the American College of Radiology Learning File. Courtesy of the ACR.

A.
motion.
B.
inadequate penetration.
C.
scattered radiation fog.
D.
excessive density.

back 110

C.
scattered radiation fog.

front 111

Which of the following is most likely to produce a radiograph with a long scale of contrast?
A.
Increased photon energy
B.
Increased screen speed
C.
Increased mAs
D.
Increased SID

back 111

A.
Increased photon energy

front 112

As grid ratio is decreased,
A.
the scale of contrast becomes longer
B.
the scale of contrast becomes shorter
C.
radiographic density decreases
D.
radiographic distortion decreases

back 112

A.
the scale of contrast becomes longer

front 113

Which of the following will contribute to the production of longer-scale radiographic contrast?
1. An increase in kV
2. An increase in grid ratio
3. An increase in photon energy

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 113

C.
1 and 3 only

front 114

A decrease from 90 to 77 kVp will result in a decrease in which of the following?

Wavelength

Scale of grays

Optical density

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 114

C.
2 and 3 only

front 115

Shape distortion is influenced by the relationship between the

x-ray tube and the part to be imaged.

body part to be imaged and the IR.

IR and the x-ray tube.

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 115

D.
1, 2, and 3

front 116

Which of the following groups of technical factors will produce the least radiographic density?
A.
400 mA, 0.010 second, 94 kV, 100-speed screens
B.
500 mA, 0.008 second, 94 kV, 200-speed screens
C.
200 mA, 0.040 second, 94 kV, 50-speed screens
D.
100 mA, 0.020 second, 80 kV, 200-speed screens

back 116

D.
100 mA, 0.020 second, 80 kV, 200-speed screens

front 117

Which of the lines indicated in Figure 4–2 represents the dynamic range offered by computed radiography/digital radiography (CR/DR)?

Courtesy FUJIFILM Medical Systems USA, Inc.

A.
Line A is representative of CR/DR.
B.
Line B is representative of CR/DR.
C.
Neither line is representative of CR/DR.
D.
Both lines are representative of CR/DR.

Figure 4–2.

back 117

A.
Line A is representative of CR/DR.

front 118

The variation in photon distribution between the anode and cathode ends of the x-ray tube is known as
A.
the line focus principle.
B.
the anode heel effect.
C.
the inverse square law.
D.
Bohr's theory.

back 118

B.
the anode heel effect.

front 119

The exposure factors of 300 mA, 0.07 second, and 95 kVp were used to produce a particular radiographic density and contrast. A similar radiograph can be produced using 500 mA, 80 kVp, and
A.
0.01 second.
B.
0.04 second.
C.
0.08 second.
D.
0.16 second.

back 119

C.
0.08 second.

front 120

The term differential absorption is related to

beam intensity

subject contrast

pathology

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 120

C.
2 and 3 only

front 121

All the following have an impact on radiographic contrast except
A.
photon energy
B.
grid ratio
C.
OID
D.
focal-spot size

back 121

D.
focal-spot size

front 122

Exposures less than the minimum response time of an AEC may be required when

using high milliamperage

using fast film–screen combinations

examining large patients or body parts

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 122

B.
1 and 2 only

front 123

If a radiograph exhibits insufficient density, this might be attributed to
1. inadequate kVp.
2. inadequate SID.
3. grid cutoff.

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 123

C.
1 and 3 only

front 124

Which of the following is (are) characteristic(s) of a 16:1 grid?
1. It absorbs a high percentage of scattered radiation.
2. It has little positioning latitude.
3. It is used with high-kVp exposures.

A.
1 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 124

D.
1, 2, and 3

front 125

A 3-inch object to be radiographed at a 36-inch SID lies 4 inches from the image recorder. What will be the image width?
A.
2.6 inches
B.
3.3 inches
C.
26 inches
D.
33 inches

back 125

B.
3.3 inches

front 126

An overall image density arising from factors other than the light or radiation used to expose the image is called
A.
fog.
B.
log-relative exposure.
C.
optical density.
D.
artifact.

back 126

A.
fog.

front 127

How would the introduction of a 6-in. OID affect image contrast?
A.
Contrast would be increased.
B.
Contrast would be decreased.
C.
Contrast would not change.
D.
The scale of contrast would not change.

back 127

A.
Contrast would be increased.

front 128

A grid is usually employed
1. when radiographing a large or dense body part.
2. when using high kilovoltage.
3. when less patient dose is required.

A.
1 only
B.
3 only
C.
1 and 2 only
D.
1, 2, and 3

back 128

C.
1 and 2 only

front 129

The interaction between x-ray photons and matter illustrated in the figure below is most likely to occur
1. in structures having a high atomic number.

Reproduced with permission from Saia DA. Radiography: Program Review and Examination Preparation, 2nd ed. Stamford, CT: Appleton & Lange, 1999.
A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3
2. during radiographic examination of the abdomen.
3. using high kV and low mAs exposure factors.

back 129

C.
2 and 3 only

front 130

Why is a very short exposure time essential in chest radiography?
A.
To avoid excessive focal-spot blur
B.
To maintain short-scale contrast
C.
To minimize involuntary motion
D.
To minimize patient discomfort

back 130

C.
To minimize involuntary motion

front 131

Recorded detail is directly related to
1. source-image distance (SID).
2. tube current.
3. focal spot size.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 131

A.
1 only

front 132

If a radiograph exposed using a 12:1 ratio grid exhibits a loss of density at its lateral edges, it is probably because the
A.
SID was too great
B.
grid failed to move during the exposure
C.
x-ray tube was angled in the direction of the lead strips
D.
central ray was off-center

back 132

A.
SID was too great

front 133

Which of the following groups of exposure factors will produce the shortest scale of contrast?
A.
200 mA, 0.25 s, 70 kVp, 12:1 grid
B.
500 mA, 0.10 s, 90 kVp, 8:1 grid
C.
400 mA, 0.125 s, 80 kVp, 12:1 grid
D.
300 mA, 0.16 s, 70 kVp, 8:1 grid

back 133

A.
200 mA, 0.25 s, 70 kVp, 12:1 grid

front 134

Changes in milliampere-seconds can affect all the following except
A.
quantity of x-ray photons produced
B.
exposure rate
C.
optical density
D.
recorded detail

back 134

D.
recorded detail

front 135

Which of the following combinations is most likely to be associated with quantum mottle?
A.
Decreased milliampere-seconds, decreased SID, fast-speed screens
B.
Increased milliampere-seconds, decreased kilovoltage, slow-speed screens
C.
Decreased milliampere-seconds, increased kilovoltage, fast-speed screens
D.
Increased milliampere-seconds, increased SID, fast-speed screens

back 135

C.
Decreased milliampere-seconds, increased kilovoltage, fast-speed screens

front 136

An increase in kilovoltage will serve to
A.
produce a longer scale of contrast
B.
produce a shorter scale of contrast
C.
decrease the radiographic density
D.
decrease the production of scattered radiation

back 136

A.
produce a longer scale of contrast

front 137

Which of the following conditions would require an increase in exposure factors?
1. Congestive heart failure
2. Pleural effusion
3. Emphysema

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 137

B.
1 and 2 only

front 138

Compared to a low ratio grid, a high ratio grid will
1. absorb more primary radiation.
2. absorb more scattered radiation.
3. allow more centering latitude.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 138

B.
1 and 2 only

front 139

The quantity of scattered radiation reaching the IR can be reduced through the use of

a fast imaging system

an air gap

a stationary grid

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 139

C.
2 and 3 only

front 140

How are mAs and radiographic density related in the process of image formation?
A.
mAs and radiographic density are inversely proportional
B.
mAs and radiographic density are directly proportional
C.
mAs and radiographic density are related to image unsharpness
D.
mAs and radiographic density are unrelated

back 140

B.
mAs and radiographic density are directly proportional

front 141

During film–screen imaging, the manifest image is formed
A.
on exposure of the film emulsion
B.
in the developer solution
C.
in the first half of the fixer process
D.
in the second half of the fixer process

back 141

B.
in the developer solution

front 142

Which of the following groups of exposure factors will produce the most radiographic density?
A.
100 mA, 50 ms
B.
200 mA, 40 ms
C.
400 mA, 70 ms
D.
600 mA, 30 ms

back 142

C.
400 mA, 70 ms

front 143

The absorption of useful radiation by a grid is called
A.
grid selectivity.
B.
grid cleanup.
C.
grid cutoff.
D.
latitude.

back 143

C.
grid cutoff

front 144

Geometric unsharpness is directly influenced by

OID

SOD

SID

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 144

A.
1 only

front 145

The appearance of underexposure on an image created using a high-speed film–screen system can be caused by all the following except
A.
insufficient mAs
B.
insufficient kV
C.
insufficient SID
D.
insufficient development

back 145

C.
insufficient SID

front 146

To be suitable for use in intensifying screens, a phosphor should have which of the following characteristics?

High conversion efficiency

High x-ray absorption

High atomic number

A.
1 only
B.
3 only
C.
1 and 2 only
D.
1, 2, and 3

back 146

D.
1, 2, and 3

front 147

Which of the following pathologic conditions are considered additive conditions with respect to selection of exposure factors?

Osteoma

Bronchiectasis

Pneumonia

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 147

D.
1, 2, and 3

front 148

Using a 48-in. SID, how much OID must be introduced to magnify an object two times?
A.
8-in. OID
B.
12-in. OID
C.
16-in. OID
D.
24-in. OID

back 148

D.
24-in. OID

front 149

Exposure rate will decrease with an increase in

SID

kilovoltage

focal-spot size

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 149

A.
1 only

front 150

An increase in kilovoltage will have which of the following effects?

More scattered radiation will be produced.

The exposure rate will increase.

Radiographic contrast will increase.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 150

B.
1 and 2 only

front 151

For which of the following examinations might the use of a grid not be necessary in an adult patient?
A.
Hip
B.
Knee
C.
Abdomen
D.
Lumbar spine

back 151

B.
Knee

front 152

Which of the following groups of technical factors would be most appropriate for the radiographic examination shown in Figure 4–30?

Figure 4–30.

Courtesy of Stamford Hospital, Department of Radiology.

A.
400 mA, 1/30 s, 72 kV
B.
300 mA, 1/50 s, 82 kV
C.
300 mA, 1/120 s, 94 kV
D.
50 mA, 1/4 s, 72 kV

back 152

A.
400 mA, 1/30 s, 72 kV

front 153

The primary source of scattered radiation is the
A.
patient.
B.
tabletop.
C.
x-ray tube.
D.
grid.

back 153

A.
patient.

front 154

The best way to control voluntary motion is
A.
immobilization of the part.
B.
careful explanation of the procedure.
C.
short exposure time.
D.
physical restraint.

back 154

B.
careful explanation of the procedure.

front 155

A particular milliampere-seconds value, regardless of the combination of milliamperes and time, will reproduce the same radiographic density. This is a statement of the
A.
line-focus principle
B.
inverse-square law
C.
reciprocity law
D.
law of conservation of energy

back 155

C.
reciprocity law

front 156

The radiographic accessory used to measure the thickness of body parts in order to determine optimal selection of exposure factors is the
A.
fulcrum
B.
caliper
C.
densitometer
D.
ruler

back 156

B.
caliper

front 157

An exposure was made at a 36-in. SID using 12 mAs and 75 kVp with a 400-speed imaging system and an 8:1 grid. A second radiograph is requested with improved recorded detail. Which of the following groups of technical factors will best accomplish this task?
A.
15 mAs, 12:1 grid, 75 kVp, 400-speed system, 36-in. SID
B.
15 mAs, 12:1 grid, 75 kVp, 400-speed system, 40-in. SID
C.
30 mAs, 12:1 grid, 75 kVp, 200-speed system, 40-in. SID
D.
12 mAs, 8:1 grid, 86 kVp, 200-speed system, 36-in. SID

back 157

C.
30 mAs, 12:1 grid, 75 kVp, 200-speed system, 40-in. SID

front 158

Which of the following has the greatest effect on radiographic density?
A.
Aluminum filtration
B.
Kilovoltage
C.
SID
D.
Scattered radiation

back 158

C.
SID

front 159

To produce a just perceptible increase in radiographic density, the radiographer must increase the
A.
mAs by 30%
B.
mAs by 15%
C.
kV by 15%
D.
kV by 30%

back 159

A.
mAs by 30%

front 160

How is SID related to exposure rate and radiographic density?
A.
As SID increases, exposure rate increases and radiographic density increases.
B.
As SID increases, exposure rate increases and radiographic density decreases.
C.
As SID increases, exposure rate decreases and radiographic density increases.
D.
As SID increases, exposure rate decreases and radiographic density decreases.

back 160

D.
As SID increases, exposure rate decreases and radiographic density decreases.

front 161

All the following are related to recorded detail except
A.
motion
B.
focal-spot size
C.
OID
D.
scattered radiation

back 161

D.
scattered radiation

front 162

The radiographic accessory used to measure the thickness of body parts in order to determine optimum selection of exposure factors is the
A.
fulcrum.
B.
caliper.
C.
densitometer.
D.
ruler.

back 162

B.
caliper.

front 163

Which of the following will result if developer replenishment is inadequate?
A.
Images with excessively high contrast
B.
Images with excessively low contrast
C.
Images with excessively high density
D.
Dry, brittle films

back 163

B.
Images with excessively low contrast

front 164

A focal-spot size of 0.3 mm or smaller is essential for
A.
small-bone radiography
B.
magnification radiography
C.
tomography
D.
fluoroscopy

back 164

B.
magnification radiography

front 165

Exposure factors of 80 kVp and 8 mAs are used for a particular nongrid exposure. What should be the new milliampere-seconds value if an 8:1 grid is added?
A.
16 mAs
B.
24 mAs
C.
32 mAs
D.
40 mAs

back 165

C.
32 mAs

front 166

All the following affect the exposure rate of the primary beam except
A.
milliamperage
B.
kilovoltage
C.
distance
D.
field size

back 166

D.
field size

front 167

Decreasing field size from 14 x 17 in. to 8 x 10 in., with no other changes, will
A.
decrease radiographic density and decrease the amount of scattered radiation generated within the part
B.
decrease radiographic density and increase the amount of scattered radiation generated within the part
C.
increase radiographic density and increase the amount of scattered radiation generated within the part
D.
increase radiographic density and decrease the amount of scattered radiation generated within the part

back 167

A.
decrease radiographic density and decrease the amount of scattered radiation generated within the part

front 168

Which of the following contribute to the radiographic contrast present in the finished radiograph?

Tissue density

Pathology

Muscle development

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 168

D.
1, 2, and 3

front 169

Compared with a low-ratio grid, a high-ratio grid will

allow more centering latitude

absorb more scattered radiation

absorb more primary radiation

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 169

C.
2 and 3 only

front 170

A grid usually is employed in which of the following circumstances?

When radiographing a large or dense body part

When using high kilovoltage

When a lower patient dose is required

A.
1 only
B.
3 only
C.
1 and 2 only
D.
1, 2, and 3

back 170

C.
1 and 2 only

front 171

A graphic diagram of signal values representing various densities within the part being imaged is called a
A.
processing algorithm
B.
DICOM
C.
histogram
D.
window

back 171

C.
histogram

front 172

Which of the following is (are) directly related to photon energy?

Kilovoltage

Milliamperes

Wavelength

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 172

A.
1 only

front 173

The CR should be directed to the center of the part of greatest interest to avoid
A.
rotation distortion
B.
magnification
C.
foreshortening
D.
elongation

back 173

A.
rotation distortion

front 174

Which of the following matrix sizes is most likely to produce the best image resolution?
A.
128 x 128
B.
512 x 512
C.
1,024 x 1,024
D.
2,048 x 2,048

back 174

D.
2,048 x 2,048

front 175

A shoulder was imaged using 300 mA, 7 ms, 70 kVp, 40-inch SID, 1.2-mm focal spot, and 100 speed screen/film combination. Which of the following changes, made to compensate for changes in optical density, would result in decreased production/visualization of blur?

1. Use of a 0.6-mm focal spot

2. Use of a 50-inch SID

3. Screens with higher intensification factor
A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 175

B.
1 and 2 only

front 176

If 84 kV and 8 mAs were used for a particular abdominal exposure with single-phase equipment, what milliampere-seconds value would be required to produce a similar radiograph with three-phase, 12-pulse equipment?
A.
24 mAs
B.
16 mAs
C.
8 mAs
D.
4 mAs

back 176

D.
4 mAs

front 177

Of the following groups of exposure factors, which will produce the most radiographic density?
A.
400 mA, 30 ms, 72-in. SID
B.
200 mA, 30 ms, 36-in. SID
C.
200 mA, 60 ms, 36-in. SID
D.
400 mA, 60 ms, 72-in. SID

back 177

C.
200 mA, 60 ms, 36-in. SID

front 178

Which of the following conditions will require an increase in x-ray photon energy/penetration?
A.
Fibrosarcoma
B.
Osteomalacia
C.
Paralytic ileus
D.
Ascites

back 178

D.
Ascites

front 179

A compensating filter is used to
A.
absorb the harmful photons that contribute only to patient dose
B.
even out widely differing tissue densities
C.
eliminate much of the scattered radiation
D.
improve fluoroscopy

back 179

B.
even out widely differing tissue densities

front 180

High-kilovoltage exposure factors are usually required for radiographic examinations using
1. water-soluble, iodinated media.
2. a negative contrast agent.
3. barium sulfate.

A.
1 only
B.
2 only
C.
3 only
D.
1 and 3 only

back 180

C.
3 only

front 181

If 400 mA, 10 ms, and 90 kV were used for a particular exposure using three-phase, 12-pulse equipment, which of the following exposure changes would be most appropriate for use on single-phase equipment to produce a similar image?
A.
Use 200 mA
B.
Use 20 mAs
C.
Use 70 kV
D.
Use 0.02 second

back 181

D.
Use 0.02 second

front 182

Underexposure of a radiograph can be caused by all the following except insufficient
A.
milliamperage (mA)
B.
exposure time
C.
Kilovoltage
D.
SID

back 182

D.
SID

front 183

Recorded detail can be improved by decreasing

the SID

the OID

patient/part motion

A.
1 only
B.
3 only
C.
2 and 3 only
D.
1, 2, and 3

back 183

C.
2 and 3 only

front 184

If 85 kV and 20 mAs were used for a particular abdominal exposure with single-phase equipment, what mAs would be required to produce a similar radiograph with 3-phase, 12-pulse equipment.
A.
40
B.
25
C.
20
D.
10

back 184

D.
10

front 185

If 82 kVp, 300 mA, and 0.05 second were used for a particular exposure using 3-phase, 12-pulse equipment, what mAs would be required, using single-phase equipment, to produce a similar radiograph?
A.
7.5
B.
20
C.
30
D.
50

back 185

C.
30

front 186

Which of the following affect(s) both the quantity and the quality of the primary beam?

Half-value layer (HVL)

Kilovoltage (kV)

Milliamperage (mA)

A.
1 only
B.
2 only
C.
1 and 2 only
D.
1, 2, and 3

back 186

C.
1 and 2 only

front 187

Methods that help to reduce the production of scattered radiation include using

compression

beam restriction

a grid

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 187

A.
1 and 2 only

front 188

If a duration of 0.05 second was selected for a particular exposure, what milliamperage would be necessary to produce 30 mAs?
A.
900
B.
600
C.
500
D.
300

back 188

B.
600

front 189

Which of the following is (are) classified as rare earth phosphors?

Lanthanum oxybromide

Gadolinium oxysulfide

Cesium iodide

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 189

B.
1 and 2 only

front 190

In comparison with 60 kV, 80 kV will

permit greater exposure latitude

produce more scattered radiation

produce shorter-scale contrast

A.
1 only
B.
2 only
C.
1 and 2 only
D.
2 and 3 only

back 190

C.
1 and 2 only

front 191

A satisfactory radiograph was made without a grid using a 72-in. SID and 8 mAs. If the distance is changed to 40 in. and an 12:1 ratio grid is added, what should be the new milliampere-seconds value?
A.
9.5 mAs
B.
12 mAs
C.
21 mAs
D.
26 mAs

back 191

B.
12 mAs

front 192

Which of the following examinations might require the use of 70 kV?

AP abdomen

Chest radiograph

Barium-filled stomach

A.
1 only
B.
2 only
C.
1 and 2 only
D.
2 and 3 only

back 192

A.
1 only

front 193

Recorded detail is directly related to

SID.

tube current.

focal-spot size.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 193

A.
1 only

front 194

Several types of exposure timers may be found on x-ray equipment. Which of the following types of timers functions to accurately duplicate radiographic densities?
A.
Synchronous
B.
Impulse
C.
Electronic
D.
Phototimer

back 194

D.
Phototimer

front 195

When involuntary motion must be considered, the exposure time may be cut in half if the kilovoltage is
A.
doubled
B.
increased by 15%
C.
increased by 25%
D.
increased by 35%

back 195

B.
increased by 15%

front 196

Greater latitude is available to the radiographer in which of the following circumstances?

Using high-kV technical factors

Using a low-ratio grid

Using low-kV technical factors

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
3 only

back 196

B.
1 and 2 only

front 197

A positive contrast agent

absorbs x-ray photons

results in a dark area on the radiograph

is composed of elements having low atomic number

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 197

A.
1 only

front 198

Which of the following quantities of filtration is most likely to be used in mammography?
A.
0.5 mm Mo
B.
1.5 mm Al
C.
1.5 mm Cu
D.
2.0 mm Cu

back 198

A.
0.5 mm Mo

front 199

An exposure was made of a part using 300 mA and 0.06 second with a 200-speed film–screen combination. An additional radiograph is requested using a 400-speed system to reduce motion unsharpness. Using 400 mA, all other factors remaining constant, what should be the new exposure time?
A.
5 ms
B.
11 ms
C.
22 ms
D.
44 ms

back 199

C.
22 ms

front 200

Because of the anode heel effect, the intensity of the x-ray beam is greatest along the
A.
path of the central ray
B.
anode end of the beam
C.
cathode end of the beam
D.
transverse axis of the IR

back 200

C.
cathode end of the beam

front 201

Which of the following groups of exposure factors will produce the longest scale of contrast?
A.
200 mA, 0.25 second, 70 kVp, 12:1 grid
B.
500 mA, 0.10 second, 90 kVp, 8:1 grid
C.
400 mA, 0.125 second, 80 kVp, 12:1 grid
D.
300 mA, 0.16 second, 70 kVp, 8:1 grid

back 201

B.
500 mA, 0.10 second, 90 kVp, 8:1 grid

front 202

In which of the following examinations should 70 kV not be exceeded?
A.
Upper GI (UGI)
B.
Barium enema (BE)
C.
Intravenous urogram (IVU)
D.
Chest

back 202

C.
Intravenous urogram (IVU)

front 203

The squeegee assembly in an automatic processor

functions to remove excess solution from films

is located near the crossover rollers

helps establish the film's rate of travel

A.
1 only
B.
2 only
C.
1 and 2 only
D.
1, 2, and 3

back 203

C.
1 and 2 only

front 204

How is source-to-image distance (SID) related to exposure rate and radiographic density?
A.
As SID increases, exposure rate increases and radiographic density increases.
B.
As SID increases, exposure rate increases and radiographic density decreases.
C.
As SID increases, exposure rate decreases and radiographic density increases.
D.
As SID increases, exposure rate decreases and radiographic density decreases.

back 204

D.
As SID increases, exposure rate decreases and radiographic density decreases.

front 205

If 300 mA has been selected for a particular exposure, what exposure time should be selected to produce 18 mAs?
A.
40 ms
B.
60 ms
C.
400 ms
D.
600 ms

back 205

B.
60 ms

front 206

The direction of electron travel in the x-ray tube is
A. filament to cathode
B. cathode to anode
C. anode to focus
D. anode to cathode

back 206

B. cathode to anode

front 207

Exposure factors of 90 kV and 3 mAs are used for a particular nongrid exposure. What should be the new milliampere-seconds (mAs) value if a 12:1 grid is added?
A. 86
B. 9
C. 12
D. 15

back 207

D. 15

front 208

The attenuation of x-ray photons is not influenced by
pathology
effective atomic number
photon quantity
A. 1 only
B. 3 only
C. 2 and 3 only
D. 1, 2, and 3

back 208

B. 3 only

front 209

The functions of automatic beam limitation devices include
reducing the production of scattered radiation
increasing the absorption of scattered radiation
changing the quality of the x-ray beam
A. 1 only
B. 2 only
C. 1 and 2 only
D. 1, 2, and 3

back 209

A. 1 only

front 210

An anteroposterior (AP) radiograph of the femur was made using 300 mA, 0.03 second, 76 kV, 40-in. SID, 1.2-mm focal spot, and a 400-speed film–screen system. With all other factors remaining constant, which of the following exposure times would be required to maintain radiographic density at a 44-in. SID using 500 mA?
A. 12 ms
B. 22 ms
C. 30 ms
D. 36 ms

back 210

B. 22 ms

front 211

If 85 kVp, 400 mA, and ? s were used for a particular exposure using single-phase equipment, which of the following milliamperage or time values would be required, all other factors being constant, to produce a similar density using three-phase, 12-pulse equipment?
A. 200 mA
B. 600 mA
C. 0.125 s
D. 0.25 s

back 211

A. 200 mA

front 212

Factor(s) that can be used to regulate radiographic density is (are)
1. milliamperage.
2. exposure time.
3. kilovoltage.

A. 1 only
B. 2 only
C. 1 and 2 only
D. 1, 2, and 3

back 212

D. 1, 2, and 3

front 213

Which of the following groups of exposure factors would be most appropriate for a sthenic adult IVU?
A. 300 mA, 0.02 s, 72 kVp
B. 300 mA, 0.01 s, 82 kVp
C. 150 mA, 0.01 s, 94 kVp
D. 100 mA, 0.03 s, 82 kVp

back 213

A. 300 mA, 0.02 s, 72 kVp

front 214

The use of which of the following is (are) essential in magnification radiography?
High-ratio grid
Fractional focal spot
Direct exposure technique
A. 1 onlyB. 2 onlyC. 1 and 3 onlyD. 1, 2, and 3

back 214

B. 2 only

front 215

Which of the following are methods of limiting the production of scattered radiation?
Using moderate ratio grids
Using the prone position for abdominal examinations
Restricting the field size to the smallest practical size
A. 1 and 2 only
B. 1 and 3 only
C. 2 and 3 only
D. 1, 2, and 3

back 215

C. 2 and 3 only

front 216

In a PA projection of the chest being used for cardiac evaluation, the heart measures 15.2 cm between its widest points. If the magnification factor is known to be 1.3, what is the actual diameter of the heart?
A. 9.7 cmB. 11.7 cmC. 19.7 cmD. 20.3 cm

back 216

B. 11.7 cm

front 217

Which of the following can affect histogram appearance?
Centering accuracy
Positioning accuracy
Processing algorithm accuracy
A. 1 onlyB. 1 and 2 onlyC. 2 and 3 onlyD. 1, 2, and 3

back 217

D. 1, 2, and 3

front 218

In digital imaging, as the size of the image matrix increases,
FOV increases
pixel size decreases
spatial resolution increases
A. 1 onlyB. 1 and 2 onlyC. 2 and 3 onlyD. 1, 2, and 3

back 218

C. 2 and 3 only

front 219

Which combination of exposure factors will most likely contribute to producing the longest-scale contrast?
Film-Screen Grid Field
mAs kVp System Ratio Size
(A) 10 70 400 5:1 14 x 17 inches
(B) 12 90 200 8:1 14 x 17 inches
(C) 15 90 200 12:1 8 x 10 inches
(D) 20 80 400 10:1 11 x 14 inches

A. Group AB. Group BC. Group CD. Group D

back 219

B. Group B

front 220

A satisfactory radiograph of the abdomen was made at a 38-in. SID using 400 mA, 60-ms exposure, and 80 kV. If the distance is changed to 42 in., what new exposure time would be required?
A. 25 msB. 50 msC. 73 msD. 93 ms

back 220

C. 73 ms

front 221

A particular radiograph was produced using 12 mAs and 85 kV with a 16:1 ratio grid. The radiograph is to be repeated using an 8:1 ratio grid. What should be the new milliampere-seconds value?
A. 3B. 6C. 8D. 10

back 221

C. 8

front 222

When the collimated field must extend past the edge of the body, allowing primary radiation to strike the tabletop, as in a lateral lumbar spine radiograph, what may be done to prevent excessive radiographic density owing to undercutting?
A. Reduce the milliampere-seconds.B. Reduce the kilovoltage.C. Use a shorter SID.D. Use lead rubber to absorb tabletop primary radiation.

back 222

D. Use lead rubber to absorb tabletop primary radiation.

front 223

A lateral radiograph of the cervical spine was made at 40 in. using 300 mA and 0.03 second exposure. If it is desired to increase the distance to 72 in., what should be the new milliampere (mA) setting, all other factors remaining constant?
A. 400 mAB. 800 mAC. 1000 mAD. 1200 mA

back 223

C. 1000 mA

front 224

Which of the following focal-spot sizes should be employed for magnification radiography?
A. 0.2 mmB. 0.6 mmC. 1.2 mmD. 2.0 mm

back 224

A. 0.2 mm

front 225

An AP radiograph of the femur was made using 300 mA, 0.03 second, 76 kV, 40-in. SID, 1.2-mm focal spot, and a 400-speed film–screen system. With all other factors remaining constant, which of the following exposure times would be required to maintain radiographic density using 87 kV, a 200-speed film–screen system, and the addition of a 12:1 grid?
A. 0.15 secondB. 0.20 secondC. 0.4 secondD. 0.6 second

back 225

A. 0.15 second

front 226

Which of the following is performed to check the correctness of the developing parameters?
A.
Densitometry
B.
A thorough cleaning of rollers
C.
A warm-up procedure
D.
Sensitometry

back 226

D.
Sensitometry

front 227

An increase in kilovoltage with appropriate compensation of milliampere-seconds will result in

increased exposure latitude.

higher contrast.

increased density.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1 and 3 only

back 227

A.
1 only

front 228

Which of the following groups of exposure factors would be most effective in eliminating prominent pulmonary vascular markings in the RAO position of the sternum?
A.
500 mA, 1/30 s, 70 kV
B.
200 mA, 0.04 second, 80 kV
C.
300 mA, 1/10 s, 80 kV
D.
25 mA, 7/10 s, 70 kV

back 228

D.
25 mA, 7/10 s, 70 kV

front 229

Which of the following factors is/are related to grid efficiency?

Grid ratio

Number of lead strips per inch

Amount of scatter transmitted through the grid

A.
1 only
B.
2 only
C.
1 and 2 only
D.
1, 2, and 3

back 229

D.
1, 2, and 3

front 230

Which one of the following is (are) used to control the production of scattered radiation?

Collimators

Optimal kV

Use of grids

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 230

B.
1 and 2 only

front 231

Types of shape distortion include

magnification

elongation

foreshortening

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 231

C.
2 and 3 only

front 232

If 300 mA has been selected for a particular exposure, what exposure time would be required to produce 6 mAs?
A.
5 ms
B.
10 ms
C.
15 ms
D.
20 ms

back 232

D.
20 ms

front 233

Central ray angulation may be required for

magnification of anatomic structures

foreshortening or self-superimposition

superimposition of overlying structures

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 233

C.
2 and 3 only

front 234

An automatic exposure control (AEC) device can operate on which of the following principles?

A photomultiplier tube charged by a fluorescent screen

A parallel-plate ionization chamber charged by x-ray photons

Motion of magnetic fields inducing current in a conductor

A.
1 only
B.
2 only
C.
1 and 2 only
D.
1, 2, and 3

back 234

C.
1 and 2 only

front 235

Exposure factors of 110 kVp and 12 mAs are used with an 8:1 grid for a particular exposure. What should be the new mAs if a 12:1 grid is substituted?
A.
3
B.
9
C.
15
D.
18

back 235

C.
15

front 236

Which of the following statements is (are) true regarding Figure 7–10?

Figure 7–10.

Courtesy of Stamford Hospital, Department of Radiology.

Excessive kilovoltage was used.

High contrast is demonstrated.

Insufficient penetration is evident.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 236

C.
2 and 3 only

front 237

Radiographic contrast is a result of

differential tissue absorption

emulsion characteristics

proper regulation of milliampere-seconds

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 237

B.
1 and 2 only

front 238

Practice(s) that enable the radiographer to reduce the exposure time required for a particular image include

use of a higher milliamperage

use of a higher kilovoltage

use of a faster film–screen combination

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 238

D.
1, 2, and 3

front 239

Decreasing field size from 14 x 17 into 8 x 10 inches will
A.
decrease radiographic density and decrease the amount of scattered radiation generated within the part.
B.
decrease radiographic density and increase the amount of scattered radiation generated within the part.
C.
increase radiographic density and increase the amount of scattered radiation generated within the part.
D.
increase radiographic density and decrease the amount of scattered radiation generated within the part.

back 239

A.
decrease radiographic density and decrease the amount of scattered radiation generated within the part.

front 240

Which of the following may be used to reduce the effect of scattered radiation on a finished radiograph?

Grids

Collimators

Compression bands

A.
1 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 240

D.
1, 2, and 3

front 241

Cassette-front material can be made of which of the following?

Carbon fiber

Magnesium

Lead

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 241

B.
1 and 2 only

front 242

A technique chart should include which of the following information?

Recommended SID

Grid ratio

Screen–film combination

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 242

D.
1, 2, and 3

front 243

If 40 mAs and a 200-speed screen–film system were used for a particular exposure, what new milliampere-seconds value would be required to produce the same density if the screen–film system were changed to 800 speed?
A.
10
B.
20
C.
80
D.
160

back 243

A.
10

front 244

The exposure factors of 400 mA, 17 ms, and 82 kV produce a milliampere-seconds value of
A.
2.35
B.
6.8
C.
23.5
D.
68

back 244

B.
6.8

front 245

The advantage(s) of high-kilovoltage chest radiography is (are) that

exposure latitude is increased

it produces long-scale contrast

it reduces patient dose

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 245

D.
1, 2, and 3

front 246

A radiograph made using 300 mA, 0.1 second, and 75 kV exhibits motion unsharpness but otherwise satisfactory technical quality. The radiograph will be repeated using a shorter exposure time. Using 86 kV and 400 mA, what should be the new exposure time?
A.
25 ms
B.
37 ms
C.
50 ms
D.
75 ms

back 246

B.
37 ms

front 247

The term latitude describes

an emulsion's ability to record a range of densities

the degree of error tolerated with given exposure factors

the conversion efficiency of a given intensifying screen

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 247

B.
1 and 2 only

front 248

Of the following groups of exposure factors, which will produce the shortest scale of radiographic contrast?
A.
500 mA, 0.040 second, 70 kV
B.
100 mA, 0.100 second, 80 kV
C.
200 mA, 0.025 second, 92 kV
D.
700 mA, 0.014 second, 80 kV

back 248

A.
500 mA, 0.040 second, 70 kV

front 249

What type of x-ray imaging uses an area beam and a photostimulable phosphor as the IR?
A.
Traditional radiography
B.
Computed radiography
C.
Digital radiography
D.
Cineradiography

back 249

B.
Computed radiography

front 250

A technique chart should be prepared for each AEC x-ray unit and should contain which of the following information for each type of examination?

Photocell(s) used

Optimum kilovoltage

Backup time

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 250

D.
1, 2, and 3

front 251

X-ray photon beam attenuation is influenced by
1. tissue type.
2. subject thickness.
3. photon quality.

A.
1 only
B.
3 only
C.
2 and 3 only
D.
1, 2, and 3

back 251

D.
1, 2, and 3

front 252

A satisfactory radiograph of the abdomen was made at a 42-inch SID using 300 mA, 0.06-second exposure, and 80 kVp. If the distance is changed to 38 inches, what new exposure time would be required?
A.
0.02 second
B.
0.05 second
C.
0.12 second
D.
0.15 second

back 252

B.
0.05 second

front 253

Better resolution is obtained with
A.
high SNR.
B.
low SNR.
C.
windowing.
D.
smaller matrix.

back 253

A.
high SNR.

front 254

In radiography of a large abdomen, which of the following is (are) effective way(s) to minimize the amount of scattered radiation reaching the image receptor (IR)?

Use of close collimation

Use of low mAs

Use of a low-ratio grid

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 254

A.
1 only

front 255

With all other factors constant, as digital image matrix size increases,
1. pixel size decreases.
2. resolution increases.
3. pixel size increases.

A.
1 only
B.
2 only
C.
1 and 2 only
D.
2 and 3 only

back 255

C.
1 and 2 only

front 256

Compression of the breast during mammographic imaging improves the technical quality of the image because

geometric blurring is decreased

less scattered radiation is produced

patient motion is reduced

A.
1 only
B.
3 only
C.
2 and 3 only
D.
1, 2, and 3

back 256

D.
1, 2, and 3

front 257

An exposure was made using 8 mAs and 60 kV. If the kilovoltage was changed to 70 to obtain longer-scale contrast, what new milliampere-seconds value is required to maintain density?
A.
2
B.
4
C.
16
D.
32

back 257

B.
4

front 258

Which of the following devices is used to overcome severe variation in patient anatomy or tissue density, providing more uniform radiographic density?
A.
Compensating filter
B.
Grid
C.
Collimator
D.
Intensifying screen

back 258

A.
Compensating filter

front 259

If 0.05 second was selected for a particular exposure, what mA would be necessary to produce 15 mAs?
A.
900
B.
600
C.
500
D.
300

back 259

D.
300

front 260

Which of the following is likely to contribute to the radiographic contrast present on the finished radiograph?

Atomic number of tissues radiographed

Any pathologic processes

Degree of muscle development

A.
1 and 2 only
B.
1 and 3 only
C.
2 and 3 only
D.
1, 2, and 3

back 260

D.
1, 2, and 3

front 261

The effect described as differential absorption is

responsible for radiographic contrast

a result of attenuating characteristics of tissue

minimized by the use of a high peak kilovoltage

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 261

D.
1, 2, and 3

front 262

A decrease from 200 to 100 mA will result in a decrease in which of the following?

Wavelength

Exposure rate

Beam intensity

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 262

C.
2 and 3 only

front 263

The steeper the straight-line portion of a characteristic curve for a particular film, the

slower is the film speed

higher is the film contrast

greater is the exposure latitude

A.
1 only
B.
2 only
C.
2 and 3 only
D.
1, 2, and 3

back 263

B.
2 only

front 264

Which of the following terms/units is used to express the resolution of a diagnostic image?
A.
Line pairs per millimeter (lp/mm)
B.
Speed
C.
Latitude
D.
Kiloelectronvolts (keV)

back 264

A.
Line pairs per millimeter (lp/mm)

front 265

A satisfactory radiograph was made without a grid, using a 72-inch SID and 8 mAs. If the distance is changed to 40 inches and an 8:1 ratio grid is added, what should be the new mAs?
A.
10 mAs
B.
18 mAs
C.
20 mAs
D.
32 mAs

back 265

A.
10 mAs

front 266

Which of the following chemicals is used in the production of radiographic film emulsion?
A.
Sodium sulfite
B.
Potassium bromide
C.
Silver halide
D.
Chrome alum

back 266

C.
Silver halide

front 267

An increase in the kilovoltage applied to the x-ray tube increases the

percentage of high-energy photons produced.

beam intensity.

patient absorption.

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 267

B.
1 and 2 only

front 268

What is the single most important factor controlling size distortion?
A.
Tube, part, IR alignment
B.
IR dimensions
C.
SID
D.
OID

back 268

D.
OID

front 269

Combinations of milliamperage and exposure time that produce a particular milliampere-seconds value will produce identical radiographic density. This statement is an expression of the
A.
inverse-square law
B.
line-focus principle
C.
reciprocity law
D.
D log E curve

back 269

C.
reciprocity law

front 270

An increase in the kilovoltage applied to the x-ray tube increases the

x-ray wavelength

exposure rate

patient absorption

A.
1 only
B.
2 only
C.
2 and 3 only
D.
1, 2, and 3

back 270

B.
2 only

front 271

The major function of filtration is to reduce
A.
image noise.
B.
scattered radiation.
C.
operator dose.
D.
patient dose.

back 271

D.
patient dose.

front 272

Which of the following methods can be used effectively to decrease differential absorption, providing a longer scale of contrast in the diagnostic range?

Using high peak kilovoltage and low milliampere-seconds factors

Using compensating filtration

Using factors that increase the photoelectric effect

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 272

B.
1 and 2 only

front 273

Which of the following could be used to improve recorded detail?

1. Slower screen/film combination

2. Smaller focal-spot size

3. Correct photocell selection
A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 273

B.
1 and 2 only

front 274

Which of the following is most likely to occur as a result of using a 30-in. SID with a 14 x 17 in. IR to radiograph a fairly homogeneous structure?
A.
Production of quantum mottle
B.
Density variation between opposite ends of the IR
C.
Production of scatter radiation fog
D.
Excessively short-scale contrast

back 274

B.
Density variation between opposite ends of the IR

front 275

Which of the following combinations will result in the most scattered radiation reaching the image receptor?
A.
Using more mAs and compressing the part
B.
Using more mAs and a higher ratio grid
C.
Using fewer mAs and more kVp
D.
Using more mAs and less kVp

back 275

C.
Using fewer mAs and more kVp

front 276

Which of the following is most likely to produce a high-quality image?
A.
Small image matrix
B.
High signal-to-noise ratio (SNR)
C.
Large pixel size
D.
Low resolution

back 276

B.
High signal-to-noise ratio (SNR)

front 277

Distortion can be caused by

tube angle

the position of the organ or structure within the body

the radiographic positioning of the part

A.
1 only
B.
1 and 2 only
C.
2 and 3 only
D.
1, 2, and 3

back 277

D.
1, 2, and 3

front 278

Which of the following is most likely to result from the introduction of a grid to a particular radiographic examination?
A.
Increased patient dose and increased contrast
B.
Decreased patient dose and decreased contrast
C.
Increased patient dose and decreased contrast
D.
Decreased patient dose and increased contrast

back 278

A.
Increased patient dose and increased contrast

front 279

Which of the following can affect radiographic contrast?
1. Processing
2. Pathology
3. OID

A.
1 only
B.
1 and 2 only
C.
1 and 3 only
D.
1, 2, and 3

back 279

D.
1, 2, and 3

front 280

All other factors remaining the same, if a 14 x 17 in. field is collimated to a 4-in.-square field, the radiographic image will demonstrate
A.
more density
B.
less density
C.
more detail
D.
less detail

back 280

B.
less density

front 281

http://radrevieweasy.com/loadBinary.aspx?name=saia8&filename=saia8_c007f133b.jpg

Both radiographic images shown in Figure 7–14 were made of the same subject using identical exposure factors. Which of the following statements correctly describe(s) these images?

From the American College of Radiology Learning File. Courtesy of the ACR.

Image A demonstrates less optical density because a shorter SID was used.

Image A demonstrates more optical density because the subject was turned PA.

Image B demonstrates more optical density because a shorter SID was used.

A.
1 only
B.
2 only
C.
3 only
D.
1 and 2 only

Figure 7–14.

back 281

C.
3 only

front 282

Exposure factors of 100 kVp and 6 mAs are used with a 6:1 grid for a particular exposure. What should be the new milliampere-seconds value if a 12:1 grid is substituted?
A.
7.5 mAs
B.
10 mAs
C.
13 mAs
D.
18 mAs

back 282

B.
10 mAs