front 1 When the speed of screen-film systems doubles, for example, when changing from a 200-speed to a 400-speed system, patient radiation exposure is _____ by approximately _____%.
| back 1 reduced; 50 |
front 2 CR imaging has greater _______ flexibility than does conventional screen-film radiography.
| back 2 kilovoltage |
front 3 In standard image intensification fluoroscopy, an x-ray beam half-value-layer (HVL) of 3- to 4-mm aluminum is considered acceptable when peak kilovoltage ranges from:
| back 3 80 to 100 |
front 4 When settings are changed from one mA to a neighboring mA station, the most that linearity can vary is _____%.
| back 4 10 |
front 5 When performing a mobile fluoroscopic procedure, to reduce the radiation exposure to the patient, the radiographer must use a minimal source-skin distance of ____ cm (____ inches).
| back 5 30; 12 |
front 6 The control panel, where technical exposure factors are selected and seen on indicators by the equipment operator, must be located:
| back 6 behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure |
front 7 To visualize smaller and lower-contrast objects during interventional procedures, high-level-control fluoroscopy uses exposure rates that are _______ those normally used in routine fluoroscopy.
| back 7 higher than |
front 8 During a fluoroscopic examination a resettable cumulative timing device times the x-ray beam-on time and sounds an audible alarm or temporarily interrupts the exposure after the fluoroscope has been activated for what length of time?
| back 8 5 minutes |
front 9 When using digital fluoroscopy systems, making use of the last-image-hold feature can:
| back 9 be an effective dose reduction technique |
front 10 Sharper size restriction of the radiographic beam is achieved when the cone or cylinder is:
| back 10 longer |
front 11 What is the purpose of radiographic beam filtration?
| back 11 To increase beam hardness, thereby reducing patient skin dose and the dose to superficial tissues |
front 12 Depending on the area of the body being examined with a fluoroscopic image intensification system, for adult patients a range of ______ kVp is generally used.
| back 12 75 to 110 |
front 13 When an exposed computed radiography imaging plate is ready to be processed, an imaging reading unit is used to scan the photostimulable phosphor imaging plate with a helium-neon laser beam. This results in the emission of violet light that is changed into an electronic signal by a device called a:
| back 13 photomultiplier tube |
front 14 Of the following procedures, which involve extended fluoroscopic time?
| back 14 1, 2, and 3 |
front 15 When compared with the traditional material in the front of a cassette (aluminum or cardboard), the cassette front containing the carbon fiber absorbs approximately _____ as much radiation, resulting in a _____ dose for the patient because _____ radiographic techniques are required to produce the recorded image.
| back 15 half; lower; lower |
front 16 Digital radiography images can be accessed:
| back 16 at several workstations at the same time, making image viewing very convenient for physicians providing patient care |
front 17 Luminance is determined by measuring the concentration of light:
| back 17 over a particular field of view |
front 18 A radiographer uses a high-speed, rare-earth screen-film system with optimal technical exposure factors to obtain posteroanterior and lateral radiographs of a 2-year-old child’s chest. To maximize reduction of radiation exposure to the patient following adequate immobilization, the radiographer should collimate the x-ray beam so that it is _______ shield the child’s reproductive organs.
| back 18 smaller than the margins of the image receptor and |
front 19 Current federal standards limit entrance skin exposure rates of general-purpose intensified fluoroscopic units to a maximum of _____ R/min (_____ ´ 2.58 ´ 10–4 C/kg/min).
| back 19 10; 10 |
front 20 Computed radiography, because of its higher exposure latitude, makes grid use on the pediatric population:
| back 20 less necessary than was previously believed |
front 21 Which of the following materials is commonly used in the tabletop of a radiographic examination table to make the tabletop as radiolucent as possible so that it will absorb only a minimal amount of radiation, thereby reducing the patient’s radiation dose?
| back 21 Carbon fiber material |
front 22 Federal government specifications recommend a minimum total filtration of _____-mm aluminum equivalent for stationary (fixed) fluoroscopic x-ray units operating above 70 kVp.
| back 22 2.5 |
front 23 Which of the following is not an x-ray beam limitation device?
| back 23 Filter |
front 24 When a fluoroscopic image is electronically amplified by an image intensification system, which of the following benefits result?
| back 24 1, 2, and 3 |
front 25 Which of the following types of filters should be used to provide uniform density when radiographing a foot in the dorsoplantar projection?
| back 25 Wedge filter |